DOI 10.1016/j.jaccas.2026.109675
Fine Ventricular Fibrillation Therapy Assurance Algorithm Failure From Low-Amplitude Far-Field Signals: Precision Sensing Strategy.
Аннотация
BACKGROUND
The VF Therapy Assurance (VFTA) algorithm in Abbott implantable cardioverter-defibrillators (ICDs) enhances detection of fine ventricular fibrillation by utilizing far-field signals when near-field sensing is inadequate. However, this algorithm may paradoxically trigger inappropriate ICD therapy in certain electrogram phenotypes.
CASE SUMMARY: A 53-year-old man with a cardiac resynchronization therapy-defibrillator for nonischemic cardiomyopathy presented after an ICD shock during light housework. Device interrogation revealed stable supraventricular tachycardia at 160 beats/min with low-amplitude far-field signals on the Coil-Can vector, triggering VFTA activation. This collapsed all detection zones into a single therapy zone, disabled supraventricular tachycardia discriminators, and delivered an inappropriate shock after only 6 beats meeting ventricular fibrillation criteria. The sensing vector was reprogrammed from Coil-Can to Tip-Can, with no recurrent inappropriate therapy on follow-up.
DISCUSSION: Low-amplitude far-field signals can inappropriately activate VFTA, leading to zone collapse and discriminator suspension during supraventricular tachycardia, resulting in unwarranted therapy delivery.
TAKE-HOME MESSAGE: Targeted sensing vector reprogramming may mitigate VFTA-mediated inappropriate shocks while preserving the algorithm's intended benefit for fine ventricular fibrillation detection.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.