MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/jcm14010187

Медицинская библиотека · Journal of clinical medicine · 2024

A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses.

Yan Weijia, Auffarth Gerd U, Khoramnia Ramin, Łabuz Grzegorz

СтатьяПеревод пока не подготовлен

Аннотация

Objectives

This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions.

Methods

With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea. The modulation transfer function (MTF) was measured at different pupil sizes for the on- and off-axis lens positions.

Results

The IOLs exhibited varying responses to decentration up to 1 mm, showing the lowest impact in polychromatic light. The least affected was AT ELANA, with an MTF loss of 15.7% to 28.4% at 50 lp/mm across the studied conditions. It was followed by PanOptix and FineVision, with the MTF loss ranging from 19.1% to 36.0% and from 21.2% to 46.6%. AT LISA showed a more substantial reduction, i.e., 41.2% to 64.8%, but it was still lower than that of Synergy (51.1% to 78.8%). When decentration was induced at a 4.5 mm distance, its effect was more evident in conditions that were closer to each IOL's SA correction. A tilt of 5° had a lesser impact than 1 mm decentration, with the effect being more severe at 4.5 mm.

Conclusions

The off-axis position affects the optical quality of trifocal IOLs. Low- rather than high-SA-correcting trifocals perform better under misalignment. In polychromatic light, the impact of misalignment is less evident, suggesting a potential mitigating effect of chromatic aberration.

Для цитирования: Yan Weijia, Auffarth Gerd U, Khoramnia Ramin, Łabuz Grzegorz. A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses.. Journal of clinical medicine. 2024. DOI: 10.3390/jcm14010187.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.