MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/meg.0b013e32835563ab

Медицинская библиотека · European journal of gastroenterology & hepatology · 2012

QuickView in small-bowel capsule endoscopy is useful in certain clinical settings, but QuickView with Blue Mode is of no additional benefit.

Koulaouzidis Anastasios, Smirnidis Alexandros, Douglas Sarah, Plevris John N

СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Analysis of small-bowel capsule endoscopy (SBCE) is time-consuming. QuickView (QV) has been added to the RAPID software to reduce the reading times. However, its validity is still under intense review. Recently, we have shown that Blue Mode (BM) provides improvements in images for most lesion categories.

AIM: To assess the validity of QuickView with white light (QVWL) and QuickView with Blue Mode (QVBM) reading, in a group of patients who underwent SBCE in our centre, by comparing it with the standard video sequence review (used as reference) by experienced SBCE readers.

METHODS

This was a retrospective study; all SBCE (August 2008-November 2011), performed with PillCam SB, with complete small-bowel visualization were included. A clinician with previous SBCE experience, unaware of the SBCE reports, reviewed prospectively the video streams on RAPID platform using QVWL and QVBM. All SBCE had been reported previously using the standard mode; these reports were considered as the reference. There were 106 cases of obscure gastrointestinal bleeding (OGIB), 81 cases of known or suspected Crohn's disease (CD) and 10 cases of polyposis syndromes.

RESULTS

The mean small-bowel evaluation was 475 (± 270) s and 450 (± 156) s for QVWL and QVBM, respectively. In the OGIB (n=106; 21 overt/85 occult), with QVWL, 54 [P0 (28), P1 (18), P2 (8)] lesions were detected, 63 [P0 (48), P1 (13), P2 (2)] with QVBM, as compared with 98 [P0 (67), P1 (23), P2 (8)] by standard (reference) reporting. For P1+P2 lesions, the sensitivity, specificity, positive predictive value and negative predictive value for QVWL (as compared with reference reporting) were 92.3, 96.3, 96 and 92.8%, respectively. For QVBM, the above values were 91, 96, 96.2 and 90.6%, respectively. Eighty-one (n=81) patients underwent SBCE for small-bowel evaluation on the basis of a clinical history of suspected or known CD. With QVWL, 71 mucosal ulcers were detected, 68 with QVBM, as compared with 155 mucosal ulcers with reference reading. Finally, in the polyposis category with QVWL and QVBM, four polypoid lesions were detected compared with seven with standard (reference) review.

CONCLUSION

QV can be used confidently in OGIB in an urgent inpatient setting and in outpatients with occult OGIB or suspected CD. Furthermore, BM does not confer any additional advantage in the QV setting. Standard review settings should be used in all other cases.

Для цитирования: Koulaouzidis Anastasios, Smirnidis Alexandros, Douglas Sarah, Plevris John N. QuickView in small-bowel capsule endoscopy is useful in certain clinical settings, but QuickView with Blue Mode is of no additional benefit.. European journal of gastroenterology & hepatology. 2012. DOI: 10.1097/meg.0b013e32835563ab.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.