DOI 10.1016/j.cca.2026.120962

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2026

Performance evaluation of automated magnetic beads extraction method for the measurement of 23 plasma steroids using dual liquid chromatography-tandem mass spectrometry (LC-MS/MS) method.

Chen Han, Xu Gang, Zhang Mei-Juan, Ying Fu-Rong, Huang Hai-Xia, Liu He-Huan, Yang Jian-Rong

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

To evaluate the performance of magnetic beads extraction method (MGE) for quantifying plasma steroids using liquid chromatography tandem mass spectrometry (LC-MS/MS).

METHODS

The evaluation encompassed calibration linearity, limit of detection (LOD), lower limit of the measuring interval (LLMI), specificity and matrix effects, trueness and recovery, intra- and inter-day precisions, repeatability, dilution consistency and stability of 23 plasma steroids extracted by MGE method were evaluated. The 23 plasma steroids were isolated and analyzed through a single sample preparation and two injections by LC-MS/MS.

RESULTS

All 23 steroids were successfully resolved chromatographically within 12.1 min. The automated MGE method exhibited excellent linearity (R2 > 0.995 for all analytes) with coefficients of variation (CVs) ranging from 3.36% to 9.28% at LLMI and from 1.58% to 16.53% at LOD. Additionally, the CVs and average deviation for repeatability were between 0.55% and 7.11%, and - 6.57% to 10.81%, respectively. Both intra-day precision (0.88% to 8.43%) and inter-day precision (0.98% to 6.51%) satisfied the acceptance criteria. The average deviation for trueness and recovery ranged from -13.93% to 9.06%. Moreover, specificity, matrix effect, dilution consistency, and stability were distinctly identified and conformed to guideline requirements. Notably, levels of AD, T, E1, and 17-OHP were significantly elevated in patients with polycystic ovary syndrome (PCOS) (p < 0.05).

CONCLUSION

The automated MGE method demonstrates high efficiency and reliability for the simultaneous quantification of 23 plasma steroids, offering a promising solution for high-throughput analysis of steroid panels in clinical in the future.

Для цитирования: Chen Han, Xu Gang, Zhang Mei-Juan, Ying Fu-Rong, Huang Hai-Xia, Liu He-Huan, Yang Jian-Rong. Performance evaluation of automated magnetic beads extraction method for the measurement of 23 plasma steroids using dual liquid chromatography-tandem mass spectrometry (LC-MS/MS) method.. Clinica chimica acta; international journal of clinical chemistry. 2026. DOI: 10.1016/j.cca.2026.120962.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.