Influence of Single Stent Size and Tandem Stents Subject to Extrinsic Ureteral Obstruction and Stent Occlusion on Stent Failure.
Аннотация
Background and Purpose: Drainage of obstructed kidney attributable to extrinsic ureteral obstruction (EUO), required to prevent renal damage, is often achieved using Double-J ureteral stents. However, these stents fail frequently, and there is considerable debate regarding what stent size, type, and configuration offer the best option for sustained drainage. In this study, we examine the impact of stent diameter and choice of single/tandem configuration, subject to EUO and various degrees of stent occlusion, on stent failure. Materials and
Methods
Computational fluid dynamics simulations and an in vitro ureter-stent experiment enabled quantification of flow behavior in stented ureters subject to EUO and stent occlusions. Various single and tandem stents under EUO were considered. In each simulation and experiment, changes in renal pressure were monitored for different degrees of stent lumen occlusion, and onset of stent failure as well as simulated distributions of fluid flow between stent and ureter lumina were determined.
Results
For an encircling EUO that completely obstructs the ureter lumen, with or without partial stent occlusion, the choice of stent size/configuration has little effect on renal pressure. The pressure increases significantly for ∼90% stent lumen occlusion, with failure at >95% occlusion, independent of stent diameter or a tandem configuration, and with little influence of occlusion length along the stent.
Conclusions
Stent failure rate is independent of stent diameter or single/tandem configuration, for the same percentage of stent lumen occlusion, in this model. Stent failure incidence may decrease for larger diameter stents and tandem configurations, because of the larger luminal area.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.