MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11282-023-00705-7

Медицинская библиотека · Oral radiology · 2023

Monte Carlo simulations of organ and effective doses and dose-length product for dental cone-beam CT.

Terashima Shingo, Sano Junta, Osanai Minoru, Toshima Keisuke, Ohuchi Kentaro, Hosokawa Yoichiro

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The use of dental cone-beam CT (CBCT) has increased in recent years. We aimed to calculate the organ and effective doses in dental CBCT using Monte Carlo simulation (MCS) and to correlate the effective dose with the dose-length product (DLP), which is a radiation dose index.

METHODS

Organ and effective doses were calculated by MCS using the adult male and female reference phantoms of the International Commission on Radiological Protection publication 110 in a half-rotation scan of the CBCT scanner Veraviewepocs 3Df. The simulations were performed by setting nine protocols in combination with the field-of-view (FOV) and imaging region. In addition, DLPs were calculated by MCS using the virtual CT Dose Index (CTDI) and CBCT phantoms, with the same protocol.

RESULTS

The effective doses were 55 and 195 μSv at the minimum FOV of Φ40 × H40 mm and maximum FOV of Φ 80 × H80 mm, respectively. The organs with the major contribution to the effective dose were the red bone marrow (11.0‒12.8%), thyroid gland (4.0‒12.7%), salivary gland (21.8‒33.2%), and remaining tissues (35.1‒45.7%). Positive correlations were obtained between the effective dose and calculated DLP using the CTDI and CBCT phantoms.

CONCLUSIONS

Organ and effective doses for each protocol of dental CBCT could be estimated using MCS. There was a positive correlation between the effective dose and DLP, suggesting that DLP can be used to estimate the effective dose of CBCT.

Для цитирования: Terashima Shingo, Sano Junta, Osanai Minoru, Toshima Keisuke, Ohuchi Kentaro, Hosokawa Yoichiro. Monte Carlo simulations of organ and effective doses and dose-length product for dental cone-beam CT.. Oral radiology. 2023. DOI: 10.1007/s11282-023-00705-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.