PURPOSE: To demonstrate the utility of a QA program for seed localization, and to design a 3D-printed template to improve the accuracy of seed placement on custom-built 125I eye plaques for uveal brachytherapy.

METHODS AND MATERIALS: A seed localization analysis tool (SLAT) was developed in MATLAB to detect variations in seed placement relative to a treatment plan. A flexible seed placement template (3D-FSPT) was designed in CAD and printed using a Formlabs Form-3 3D printer. The 3D-FSPT and SLAT were tested using 3D-printed model eye plaques with nonradioactive seeds arranged following clinically-relevant treatment plans. Five clinical plaques were also evaluated.

RESULTS

SLAT detected submillimeter scale variations in seed position with 2.3% error relative to the plan's seed coordinates, and with an uncertainty of ± 0.01 mm. The average seed displacement on the model plaques with free-handed seed placement was 1.31 mm (SD = 0.61), and the average seed orientation difference was 5.27 degrees (SD = 4.77). The average seed displacement on the clinical plaques was 0.77 mm (SD = 0.42), and the average seed orientation difference was 4.41 degrees (SD = 3.00). For the clinical plaques, changes in dosimetry to the tumor apex and critical eye structures were within acceptable tolerances. Seed displacement (mm) (p < 0.001) and seed orientation differences (degrees) (p = 0.008) were significantly lower using the template to guide seed placement on the model plaques compared to free-handed seed placement.

CONCLUSIONS

The feasibility of a 3D-FSPT and SLAT is demonstrated for improving seed placement accuracy relative to a treatment plan.

Для цитирования: Calabresi Kaitlyn, Emrich Jacqueline, Belko Sara, Pugliese Robert, Komarnicky-Kocher Lydia, Mourtada Firas. 3D-printed template design to improve 125I seed plaque assembly accuracy for uveal brachytherapy.. Brachytherapy. 2025. DOI: 10.1016/j.brachy.2025.01.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.