OBJECTIVE

To validate the use of temperature-controlled radiofrequency energy applied to the soft palate in a multicenter setting for reduction of snoring in a minimally morbid manner.

METHODS

Prospective, nonrandomized multicenter study of 113 patients who had a respiratory disturbance index less than 15 and minimum oxygen saturation not less than 85% and who were seeking treatment for habitual disruptive snoring. Patients were given either single or multiple lesions to the soft palate during each treatment session.

RESULTS

Patients received 1978 J on average with an overall average of 2.4 treatments. Snoring scores went from an average of 7.8 (visual analog scale (VAS), 0-10) pretreatment to 3.2 posttreatment. Pain was minimal, averaging 1.7 (VAS 0-10) on days 1 to 6. Complications were few and transient, and mild.

CONCLUSIONS

The multiple lesion protocol was the most successful; reducing snoring from 7.6 to 2.7, on a VAS with an average of 1232 J delivered over 1.6 treatments. Temperature-controlled radiofrequency was found to be a minimally invasive, well-tolerated procedure that was safe and efficacious in this study group.

Для цитирования: Sher A E, Flexon P B, Hillman D, Emery B, Swieca J, Smith T L, Cartwright R, Dierks E, Nelson L. Temperature-controlled radiofrequency tissue volume reduction in the human soft palate.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2001. DOI: 10.1067/mhn.2001.119141.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.