MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00464-010-1527-7

Медицинская библиотека · Surgical endoscopy · 2010

The importance of precompression time for secure stapling with a linear stapler.

Nakayama Shinya, Hasegawa Suguru, Nagayama Satoshi, Kato Shigeru, Hida Koya, Tanaka Eiji, Itami Atsushi, Kubo Hajime, Sakai Yoshiharu

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

One prerequisite for the safe stapling of anastomoses is the formation of perfectly aligned B-shaped staples. Using an animal model, this study investigated whether precompression enhances secure staple formation.

METHODS

A 45-mm linear stapler with a blue cartridge (staple leg length, 3.5 mm) was used on three portions each of six porcine stomachs (cardia, center, and pylorus). Staple shape and height were examined according to the precompression time (0, 1, or 5 min) before firing. The shape was classified as optimal or suboptimal, and the heights of the staples were measured individually. The completeness of the proximal staple lines also was compared with the distal lines.

RESULTS

The optimal staple rate in the 5-min group (52.7%) was significantly higher than in the 1-min group (28.7%; p<0.001) or the 0-min group (17.1%; p=0.002). The optimal staple rate in the 5-min group for the cardiac portion (mean wall thickness, 2.7 mm) reached 98.9%. However, this rate was less than 50% in the center (5.3 mm) and pylorus (4.2 mm) portions regardless of the precompression time. Compared with the root side, the top side of the cartridge had a lower optimal rate (45.3 vs. 18.7%; p<0.001). A high correlation of completeness was observed between the distal and proximal sides of the stump.

CONCLUSIONS

When a linear stapling device is used, proper staple formation is correlated with intestinal wall thickness, and a sufficient amount of precompression time is effective for gaining the secure staple formation.

Для цитирования: Nakayama Shinya, Hasegawa Suguru, Nagayama Satoshi, Kato Shigeru, Hida Koya, Tanaka Eiji, Itami Atsushi, Kubo Hajime, Sakai Yoshiharu. The importance of precompression time for secure stapling with a linear stapler.. Surgical endoscopy. 2010. DOI: 10.1007/s00464-010-1527-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.