DOI 10.1016/j.cjca.2019.01.024

Медицинская библиотека · The Canadian journal of cardiology · 2019

Possible Transmitral Pressure Gradient Elevation in MitraClip XTR.

Miura Mizuki, Gavazzoni Mara, Taramasso Maurizio, Maisano Francesco

Клинический случайСтатьяVideo-Audio MediaПеревод в очереди

Аннотация

A 74-year-old man presented with dyspnea due to severe mitral regurgitation (MR) caused by rupture of the chordae tendineae. Percutaneous edge-to-edge mitral valve repair was performed using the MitraClip system (Abbott Vascular, Santa Clara, CA). After leaflet grasping via an XTR clip, transesophageal echocardiography (TEE) revealed marked reduction of MR, but elevation of the transmitral gradient was detected no matter how the position of grasping was changed. Finally, we replaced the XTR to the NTR system. After leaflet grasping via an NTR clip, TEE revealed marked reduction of MR from severe to mild and an adequate transmitral gradient.

Для цитирования: Miura Mizuki, Gavazzoni Mara, Taramasso Maurizio, Maisano Francesco. Possible Transmitral Pressure Gradient Elevation in MitraClip XTR.. The Canadian journal of cardiology. 2019. DOI: 10.1016/j.cjca.2019.01.024.