Violakis Georgios, Antonakis Pantelis, Kritsotakis Emmanouil, Kozonis Theodoros, Chardalias Leonidas, Papalois Apostolos, Agrogiannis Georgios, Kampouroglou Effrosyni, Vardakis Nikolaos, Kostakis Stylianos, Athanasaki Eleni, Zhang Zhenyu, Angelmahr Martin, Konstadoulakis Manousos, Polygerinos Panagiotis

Аннотация

A novel endoscopic palpation tool (EPT), designed for tactile and stiffness sensing using fiber Bragg gratings (FBGs) was evaluated in a surgical environment for intraoperative safety and effectiveness. The EPT consisted of four FBGs arranged in a cross pattern and embedded within an elastic, hollow, silicone hemispherical dome designed to deform upon contact with soft tissue. The EPT was employed to scan both in vivo and ex vivo tissue samples. Monitoring of porcine vital signs during minimally invasive and open surgical procedures showed no significant changes during use of the EPT. Perioperative blood tests including inflammatory markers and liver and renal function studies were unremarkable. Histopathological analyses of tissues involved (liver, spleen, bowel, and abdominal wall) showed no evidence of inflammation, necrosis, or tissue damage, confirming the device's biocompatibility. To the best of our knowledge, this is the first study reporting in vivo stiffness measurements using an FBG-based EPT. The probe successfully distinguished between soft and hard tissue regions' relative stiffness. Furthermore, successive measurements on liver samples demonstrated the device's ability to generate stiffness maps, enabling clear visualization of spatial variation in tissue stiffness.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты