Temperature and luminosity outputs of endoscopes used in transcanal endoscopic ear surgery: an experimental study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Milner T D, Jaffer M, Iyer A
Аннотация
OBJECTIVE: To establish the relationship between endoscope temperatures and luminosity with a variety of light source types, endoscope ages, endoscope sizes, angles and operative distance in transcanal endoscopic ear surgery.
METHODS: Transcanal endoscopic ear surgery was simulated in an operating theatre using 7 mm plastic suction tubing coated in insulating tape. An ATP ET-959 thermometer was used to record temperatures, and a Trotec BF06 lux meter was used to measure luminosity. Luminosity and temperature recordings were taken at 0 mm and 5 mm from the endoscope tip.
RESULTS: Thermal energy transfer from operating endoscopes is greatest when: the light intensity is high, there is a light-emitting diode light source and the endoscope is touching the surface. Additionally, larger-diameter endoscopes, angled endoscopes and new endoscopes generated greater heat.
CONCLUSION: It is recommended that operative light intensity is maintained at the lowest level possible, and that the surgeon avoids contact between patient tissues and the endoscope tip.
METHODS: Transcanal endoscopic ear surgery was simulated in an operating theatre using 7 mm plastic suction tubing coated in insulating tape. An ATP ET-959 thermometer was used to record temperatures, and a Trotec BF06 lux meter was used to measure luminosity. Luminosity and temperature recordings were taken at 0 mm and 5 mm from the endoscope tip.
RESULTS: Thermal energy transfer from operating endoscopes is greatest when: the light intensity is high, there is a light-emitting diode light source and the endoscope is touching the surface. Additionally, larger-diameter endoscopes, angled endoscopes and new endoscopes generated greater heat.
CONCLUSION: It is recommended that operative light intensity is maintained at the lowest level possible, and that the surgeon avoids contact between patient tissues and the endoscope tip.