MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.joto.2021.08.002

Медицинская библиотека · Journal of otology · 2021

FDA MAUDE database analysis of titanium middle ear prosthesis.

Carey Jaclyn, Gabbireddy Sairisheel, Mammen Luke, Rosamilia Gianna, Patel Varun, Foyt David, Parnes Steven

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Partial ossicular replacement (PORP) and total ossicular replacement prostheses (TORP) are used to restore ossicular chain function. Despite favorable auditory outcomes, these interventions have associated risks and complications. This study examines the FDA MAUDE database for ossicular chain prosthesis adverse events to highlight complications, interventions, and root cause analysis (RCA) findings.

MATERIALS AND

METHODS

The MAUDE database was searched for Medical Device Reports (MDRs) relating to PORPs and TORPs from 2010 to 2020. MDR event descriptions were reviewed, and adverse events were identified as a device issue, patient issue, and/or packaging issue that occurred intraoperatively or postoperatively.

RESULTS

Our search identified 70 MDRs which included 110 reported adverse events. Events consisted of 63 device issues, often due to device breaks and displacements, 39 patient issues, including common complaints of hearing loss and erosion, and 8 packaging issues. When comparing PORPs and TORPs, TORPs had more reported device issues whereas PORPs had more packaging issues. Intraoperative device issues were commonly resolved by completing the procedure with a backup device and most postoperative device issues required additional surgery. For devices returned to the manufacturer, RCA determined that most breaks were caused by modification and/or mishandling or that the product met specifications with an undetermined cause for the break.

CONCLUSION

Device issues were the most common adverse events and frequently required subsequent intervention. Displacement occurred more often with TORPs and was associated with changes in hearing or erosion. The findings of this study are purely descriptive and may not have direct clinical relevance.

Для цитирования: Carey Jaclyn, Gabbireddy Sairisheel, Mammen Luke, Rosamilia Gianna, Patel Varun, Foyt David, Parnes Steven. FDA MAUDE database analysis of titanium middle ear prosthesis.. Journal of otology. 2021. DOI: 10.1016/j.joto.2021.08.002.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.