MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jvoice.2013.01.006

Медицинская библиотека · Journal of voice : official journal of the Voice Foundation · 2013

Validity of a new respiratory resistance measurement device to detect glottal area change.

Gallena Sally J K, Tian Wei, Johnson Arthur T, Vossoughi Jafar, Sarles Stephen A, Solomon Nancy Pearl

Сравнительное исследованиеСтатьяПоддержка NIH: внешняя программаВалидационное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine the correlation between respiratory resistance (Rr) values measured with the Airflow Perturbation Device (APD) to laryngoscopic images of glottal area (GA) in feigned paradoxical vocal fold motion (PVFM), also known as vocal cord dysfunction.

HYPOTHESIS: There is a strong inverse relationship between Rr and GA such that laryngeal constriction can be detected and quantified by APD-measured Rr.

STUDY

DESIGN

Prospective, single subject study.

METHODS

A healthy adult feigned breathing that was characteristic of PVFM. Rr and GA were simultaneously recorded, synchronized, and analyzed for three complete breathing cycles with significant glottal constriction occurring during inspiration.

RESULTS

Cross-correlation analysis revealed a strong negative correlation (-0.824) between GA and Rr during feigned PVFM breathing such that Rr increased when GA decreased.

CONCLUSION

APD-measured Rr appears to be a viable noninvasive method for diagnostic screening and monitoring of treatment outcomes for individuals presenting with dyspnea related to PVFM.

Для цитирования: Gallena Sally J K, Tian Wei, Johnson Arthur T, Vossoughi Jafar, Sarles Stephen A, Solomon Nancy Pearl. Validity of a new respiratory resistance measurement device to detect glottal area change.. Journal of voice : official journal of the Voice Foundation. 2013. DOI: 10.1016/j.jvoice.2013.01.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.