Evaluation of the COBAS(®) AmpliPrep/COBAS(®) TaqMan(®) HCV Test, v2.0 and comparison to assays used in routine clinical practice in an international multicenter clinical trial: The ExPECT study.

Сравнительное исследованиеОценочное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kessler Harald H, Cobb Bryan R, Wedemeyer Heiner, Maasoumy Benjamin, Michel-Treil Veronique, Ceccherini-Nelli Luca, Bremer Birgit, Hübner Margit, Helander Anna, Khiri Hacene, Heilek Gabrielle, Simon Christian O, Luk Kevin, Aslam Shagufta, Halfon Philippe

Аннотация

BACKGROUND: The COBAS(®) AmpliPrep(®)/COBAS(®) TaqMan(®) HCV Test, v2.0 (CAP/CTM2) is used for HCV RNA viral load monitoring.

OBJECTIVES: The performance of the CAP/CTM2 was compared to other widely used tests, including a manual version of the assay (the COBAS(®) TaqMan(®) HCV Test, v2.0 for use with the High Pure System, HPS/CTM2) predominantly used during phase III clinical trials for the new direct acting antiviral therapies.

STUDY DESIGN: Low HCV RNA level comparisons were performed across tests (Abbott Realtime HCV Test, ART; COBAS(®) AmpliPrep(®)/COBAS(®) TaqMan(®) HCV Test, v1.0, CAP/CTM1; CAP/CTM2; and HPS/CTM2) using dilutions of the 2nd HCV WHO International Standard. Additionally, the clinical performance of the CAP/CTM2 was evaluated with 421 leftover HCV RNA-positive routine clinical samples.

RESULTS: All quantifiable WHO dilutions were within ±0.3log10IU/mL of the expected results across tests and the analytical sensitivity resulted in a limit of detection of 12IU/mL (95% confidence interval, 10, 15). When clinical samples were tested the results for 87% (367 of 421) of all sample comparisons were within ±0.5log10IU/mL. When low viral load results (25-3500IU/mL) were compared, values obtained by the ART assay were significantly lower (p<0.0001) than those obtained with the CAP/CTM2.

CONCLUSIONS: The new CAP/CTM2 showed good accuracy with comparable sensitivity to comparator assays. The new kit is well-suited for use in the routine diagnostic laboratory, especially for accurate monitoring of patients receiving triple therapy or interferone-free regimens.