Mechanism of dentin hardness measurements using a dentin hardness measuring device with a light-emitting diode.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kondo Sota, Tomioka Yutaka, Fujimoto Naohiro, Mine Atsushi, Yamaguchi Satoshi, Okumura Saeko, Tanimoto Hiroaki, Yasuo Kenzo, Yoshikawa Kazushi, Yamamoto Kazuyo, Hazama Hisanao

Аннотация

SIGNIFICANCE: Managing caries is imperative in a rapidly aging society. Current diagnoses use qualitative indices. However, a quantitative evaluation of hardness in a clinical setting may lead to more accurate diagnoses. Previously, hardness meter using indenter with light for tooth monitoring (HAMILTOM) was developed to quantitatively measure tooth hardness. Herein, the physical interpretation of dentin hardness measured using HAMILTOM and the dentin hardness measurement mechanism are discussed.

AIM: This study evaluates the mechanism of dentin hardness measurements using HAMILTOM physically and compare the invasiveness to dentin by HAMILTOM with those using a dental probe for palpation.

APPROACH: Eleven bovine dentin samples were used to create caries models. HAMILTOM measured the dark areas, and its indentations were observed using scanning electron microscopy. Also, its invasiveness was evaluated by comparing the results with those from dental probe palpation.

RESULTS: The indentation areas were smaller than the dark areas in HAMILTOM, which may be due to exuded water from the dentin sample and the elastic recovery of dentin sample. Additionally, the dental probe indentation was deeper than the HAMILTOM indentations.

CONCLUSIONS: The results demonstrate that the indentation areas were smaller than the dark areas measured by HAMILTOM, which might contain the influence of exuded water and the deformation of dentin sample. Also, HAMILTOM is less invasive than dental probe palpation. In the future, HAMILTOM may become a standard hardness measuring method to diagnose root caries.