DOI 10.12659/msm.946624

Медицинская библиотека · Medical science monitor : international medical journal of experimental and clinical research · 2024

Accuracy of 3 Intraoral Scanners in Recording Impressions for Full Arch Dental Implant-Supported Prosthesis: An In Vitro Study.

Jain Saurabh, Sayed Mohammed E, Khawaji Reem Abdullah A, Hakami Ghada Ali J, Solan Eman Hassan M, Daish Manal A, Jokhadar Hossam F, AlResayes Saad Saleh, Altoman Majed S, Alshehri Abdullah Hasan, Alqahtani Saeed M, Alamri Mohammad, Alshahrani Ahid Amer, Al-Najjar Hind Ziyad, Mattoo Khurshid

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND This study used an edentulous mandibular resin model with 6 parallel osteotomy sites and aimed to compare the accuracy (trueness and precision) of 10 digital impressions using 3 intraoral scanners, the 3Shape TRIOS 5, Medit i700, and Primescan, using Medit Link v3.3.2 software. MATERIAL AND METHODS A model simulating a patient's lower jaw was surgically prepared at 6 parallel sites (implant osteotomy), allowing placement of 6 implant analogues. Matrix-Direct transfer abutments were attached to the analogs, and a reference scan was obtained using a CeramilMap 600 extraoral scanner. Three intraoral scanners (3Shape TRIOS 5, Medit i700, and Primescan) made 10 digital impressions of each model. The data obtained were superimposed and compared using software (Medit Link 3.3.2) to evaluate accuracy. Mean values were statistically analyzed using one-way ANOVA and post hoc Tukey test. Differences were considered significant at a P value of less than 0.05. RESULTS The TRIOS 5 intraoral scanner displayed the lowest deviation for precision (37.8±4.53 μm) and trueness (54.9±11 μm), followed by Medit i700 (precision 40.6±4.17 μm, trueness 60.5±10.9 μm), whereas the highest deviation (precision: 49.1±8.31 μm, trueness: 72.3±10.4 μm) was reported when Primescan intraoral scanner was used for recording impressions of full arch implants. When the 3 intraoral scanners were compared, a statistically significant difference was observed in terms of precision (P<0.005) and trueness (P<0.005). CONCLUSIONS TRIOS 5 intraoral scanner displayed the lowest deviation values for precision and trueness (more accurate), followed by Medit i700 and Primescan intraoral scanners. However, deviation values of all scanners were within clinically acceptable limits.

Для цитирования: Jain Saurabh, Sayed Mohammed E, Khawaji Reem Abdullah A, Hakami Ghada Ali J, Solan Eman Hassan M, Daish Manal A, Jokhadar Hossam F, AlResayes Saad Saleh, Altoman Majed S, Alshehri Abdullah Hasan, Alqahtani Saeed M, Alamri Mohammad, Alshahrani Ahid Amer, Al-Najjar Hind Ziyad, Mattoo Khurshid. Accuracy of 3 Intraoral Scanners in Recording Impressions for Full Arch Dental Implant-Supported Prosthesis: An In Vitro Study.. Medical science monitor : international medical journal of experimental and clinical research. 2024. DOI: 10.12659/msm.946624.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.