DOI 10.1016/j.jaccas.2026.109927
Dual Rare Lead-Related Complications: A Case Report Highlighting Troubleshooting Challenges During ICD Lead Implantation.
Аннотация
BACKGROUND
LMNA-related dilated cardiomyopathy is associated with ventricular remodeling and a high arrhythmic risk, frequently requiring implantable cardioverter-defibrillator (ICD) implantation. Although transvenous systems are generally safe, rare lead-related complications may occur.
CASE SUMMARY: We report a 51-year-old man with LMNA-related dilated cardiomyopathy who experienced 2 rare and sequential ICD lead complications. An initially septal right-ventricular lead was complicated by myocardial transfixion because of anterior apical positioning, revealed by cardiac computed tomography after a rise in pacing thresholds. During reintervention, a second lead was implanted but was later found to have traversed an unrecognized sinus venosus atrial septal defect with partial anomalous pulmonary venous return, resulting in inadvertent left-ventricular lead placement. The lead was successfully repositioned under transesophageal echocardiographic guidance. Three-dimensional computed tomography reconstruction clearly demonstrated both abnormal lead positions.
DISCUSSION AND TAKE-HOME MESSAGE: This case highlights the importance of systematic troubleshooting, awareness of altered anatomy, and multimodality imaging during ICD implantation.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.