Quantifying false positional corrections due to facial motion using SGRT with open-face Masks.

Bry Victoria, Licon Anna Laura, McCulloch James, Kirby Neil, Myers Pamela, Saenz Daniel, Stathakis Sotirios, Papanikolaou Niko, Rasmussen Karl

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Studies have evaluated the viability of using open-face masks as an immobilization technique to treat intracranial and head and neck cancers. This method offers less stress to the patient with comparable accuracy to closed-face masks. Open-face masks permit implementation of surface guided radiation therapy (SGRT) to assist in positioning and motion management. Research suggests that changes in patient facial expressions may influence the SGRT system to generate false positional corrections. This study aims to quantify these errors produced by the SGRT system due to face motion.

METHODS

Ten human subjects were immobilized using open-face masks. Four discrete SGRT regions of interest (ROIs) were analyzed based on anatomical features to simulate different mask openings. The largest ROI was lateral to the cheeks, superior to the eyebrows, and inferior to the mouth. The smallest ROI included only the eyes and bridge of the nose. Subjects were asked to open and close their eyes and simulate fear and annoyance and peak isocenter shifts were recorded. This was performed in both standard and SRS specific resolutions with the C-RAD Catalyst HD system.

RESULTS

All four ROIs analyzed in SRS and Standard resolutions demonstrated an average deviation of 0.3 ± 0.3 mm for eyes closed and 0.4 ± 0.4 mm shift for eyes open, and 0.3 ± 0.3 mm for eyes closed and 0.8 ± 0.9 mm shift for eyes open. The average deviation observed due to changing facial expressions was 1.4 ± 0.9 mm for SRS specific and 1.6 ± 1.6 mm for standard resolution.

CONCLUSION

The SGRT system can generate false positional corrections for face motion and this is amplified at lower resolutions and smaller ROIs. These errors should be considered in the overall tolerances and treatment plan when using open-face masks with SGRT and may warrant additional radiographic imaging.

Для цитирования: Bry Victoria, Licon Anna Laura, McCulloch James, Kirby Neil, Myers Pamela, Saenz Daniel, Stathakis Sotirios, Papanikolaou Niko, Rasmussen Karl. Quantifying false positional corrections due to facial motion using SGRT with open-face Masks.. Journal of applied clinical medical physics. 2021. DOI: 10.1002/acm2.13170.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.