Evaluating CT Dose Variation Across Scanner Technologies: Implications for Compliance with New CMS CT Radiation Dose Measure.

Rehani Madan M, Mataac Maria T, Li Xinhua

СтатьяПеревод пока не подготовлен

Аннотация

RATIONALE AND

OBJECTIVES

In 2025, the Centers for Medicare and Medicaid Services introduced a computed tomography (CT) dose measure for pay-for-performance programs. Hospitals employ diverse scanner fleets, but the impact of scanner technologies on dose benchmarking remains unclear. This study evaluates dose variation across scanner models and its benchmarking implications.

MATERIALS AND

METHODS

A retrospective analysis examined CT exams from January to December 2023 at a quaternary-care hospital, focusing on median-sized adults (water-equivalent diameter: 16-19cm head, 18-22cm neck, 29-33cm torso). Dose indices from seven scanner models across eight adult exams were evaluated. The 50th and 75th percentile doses were calculated per exam and scanner model combination, compared to American College of Radiology achievable doses and diagnostic reference levels.

RESULTS

Analyzing 34,166 studies, significant dose variations with scanner models emerged. Head without contrast (N=21,654) had median volume CT-dose-index (CTDIvol) of 36.1-68.3mGy and dose-length-product (DLP) 704-1307.8mGy·cm; 75th percentiles were 43.1-69.1mGy and 838.2-1378.1mGy·cm. Chest with contrast (N=3065) showed median CTDIvol of 6.7-16.1mGy and DLP 263.8-579.7mGy·cm; 75th percentiles were 8.2-19.5mGy and 329-713.7mGy·cm. Abdomen/pelvis with contrast (N=2740) had median CTDIvol of 8.8-15.2mGy and DLP 429.3-782.1mGy·cm; 75th percentiles were 10-18.5mGy and 533.4-941.5mGy·cm. While the number of studies was smaller, five other exams also showed large dose variations across scanner models.

CONCLUSION

Single-value dose benchmarks ignoring scanner technology may be inadequate, even for similar-sized patients, potentially requiring scanner removal. Incorporating benchmarks with diverse technologies could prevent increased healthcare costs and patient care disruptions.

Для цитирования: Rehani Madan M, Mataac Maria T, Li Xinhua. Evaluating CT Dose Variation Across Scanner Technologies: Implications for Compliance with New CMS CT Radiation Dose Measure.. Academic radiology. 2025. DOI: 10.1016/j.acra.2025.03.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.