Automated liver stiffness measurements with magnetic resonance elastography.
Аннотация
PURPOSE: To provide a fully automated algorithm for obtaining stiffness measurements from hepatic magnetic resonance elastography (MRE) images that are consistent with measurements performed by expert readers.
MATERIALS AND
METHODS
An initial liver contour was found using an adaptive threshold and expanded using an active contour to select a homogeneous area of the liver. The confidence map generated during the stiffness calculation was used to select a region of reliable wave propagation. The average stiffness within the automatically generated region of interest (ROI) was compared to measurements by two trained readers in a set of 88 clinical test cases ranging from healthy to severely fibrotic.
RESULTS
The stiffness measurements reported by the readers differed by -6.76% ± 22.8% (95% confidence) and had an intraclass correlation coefficient (ICC) of 0.972 (P < 0.05). The algorithm and the more experienced reader differed by 4.32% ± 14.9 with an ICC of 0.987.
CONCLUSION
The automated algorithm performed reliably, even though MRE acquisitions often have motion artifacts present. The correlation between the automated measurements and those from the trained readers was superior to the correlation between the readers.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.