DOI 10.1016/j.bja.2022.07.015

Медицинская библиотека · British journal of anaesthesia · 2022

Comparison of two sizes of GlideScope® blades in tracheal intubation of infants: a randomised clinicaltrial.

Kwon Ji-Hye, Chung Yoon Joo, Her Sukyoung, Jeong Ji Seon, Kim Chungsu, Min Jeong-Jin

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The appropriate size of GlideScope® blade for tracheal intubation in neonates and premature infants has not been established. We evaluated the impact of the size of the GlideScope® blade on the time taken for intubation in infants weighing 2.5-3.6 kg.

METHODS

Sixty infants weighing 2.5-3.6 kg were randomly assigned to use of the size 1 blade (n=30) or the size 2 blade (n=30). The primary outcome was the time taken to intubate. Components related to the laryngoscopic view which could affect the duration of the intubating process were also analysed.

RESULTS

The time required for tracheal intubation was shorter with the size 2 blade than with the size 1 blade (16 [14-20] s vs 22 [18-25] s, P=0.002; median difference=-5; 95% confidence interval, -7 to -2). The rate at which the tip of the tracheal tube was located at the centre of the laryngeal inlet was higher with the size 2 blade than with the size 1 blade (83% vs 40%, P<0.001). Correlation analysis indicated that the time required to find the tip of the tube was related to how far the lower border of the arytenoid cartilages was located from the mid-horizontal line of the monitor (r=0.28, P=0.033).

CONCLUSION

Use of the size 2 blade significantly reduced the time required to intubate the trachea, compared with the size 1 GlideScope® blade in infants.

CLINICAL TRIAL REGISTRATION: KCT 0003867.

Для цитирования: Kwon Ji-Hye, Chung Yoon Joo, Her Sukyoung, Jeong Ji Seon, Kim Chungsu, Min Jeong-Jin. Comparison of two sizes of GlideScope® blades in tracheal intubation of infants: a randomised clinicaltrial.. British journal of anaesthesia. 2022. DOI: 10.1016/j.bja.2022.07.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.