MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000001661

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2018

A Prospective Study of Pain From Magnetic Resonance Imaging With Cochlear Implant Magnets In Situ.

Pross Seth E, Ward Bryan K, Sharon Jeffrey D, Weinreich Heather M, Aygun Nafi, Francis Howard W

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To describe changes in pain associated with magnetic resonance imaging (MRI) with cochlear implant magnets in place.

STUDY

DESIGN

Prospective, single-arm study.

SETTING

Tertiary referral center.

SUBJECTS: Adults with cochlear implants requiring MRI.

INTERVENTION: Tight head wrapping over internal device during MRI.

MAIN OUTCOME MEASURES: Change in pain score using an 11-point visual analogue scale, duration/completion of MRI, body mass index (BMI), quality of pain, status of the skin, functioning of implant, displacement/polarity change of magnet, willingness to repeat MRI without magnet removal.

RESULTS

A total of 27 subjects obtained 42 MRI scans. Subjects were 59% male with age range of 21 to 80 years. All three manufacturers were represented. Forty-eight percent of scans imaged the brain/head while 52% imaged other sites. The mean individual change in pain was 3.9 (SD 3.5, range 0-10). The pain was most commonly described as "pressure", "heat", or "sharp". There was no significant correlation between change in pain and scan duration, BMI, or body part imaged. Eighty-eight percent of the scans were completed. There were no skin complications except temporary erythema (29%) and there were no magnet/device complications. Eighty-eight percent said they would undergo MRI without magnet removal again.

CONCLUSION

The pain associated with obtaining an MRI without cochlear implant magnet removal is highly variable. Increase in pain is not related to duration of MRI scan, body part imaged, or BMI. Despite the pain, almost all patients prefer MRI scanning with the magnet in place, to avoid two surgical procedures.

Для цитирования: Pross Seth E, Ward Bryan K, Sharon Jeffrey D, Weinreich Heather M, Aygun Nafi, Francis Howard W. A Prospective Study of Pain From Magnetic Resonance Imaging With Cochlear Implant Magnets In Situ.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2018. DOI: 10.1097/mao.0000000000001661.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.