DOI 10.1007/s10840-026-02430-7

Медицинская библиотека · Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026

Efficacy of closed loop stimulation in left bundle branch area pacing.

Yarmohammadi Hirad, Salehi Bobak, Shea Margaret, Harrell Camden, Miller Crystal, Hayes David

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Closed-Loop Stimulation (CLS) adjusts pacing rate based on beat-to-beat impedance changes reflecting myocardial contractility and autonomic tone. While early studies focused on right ventricular apical leads, conduction system pacing, particularly left bundle branch area pacing (LBBAP) offers a more physiologic approach. The effect of septal lead position on CLS performance remains unclear.

OBJECTIVE

To investigate the impact of LBBAP versus alternative right ventricular lead positions on CLS performance.

METHODS

We conducted a retrospective cohort study using the CERTITUDE registry, which includes longitudinal remote monitoring data from over 55,000 U.S. patients with Biotronik devices. Patients with dual-chamber pacemakers implanted starting January 1, 2021, with ≥ 1 year of follow-up were included, excluding those who were > 110 years old at implant, with ≥ 20% atrial fibrillation burden, < 50% transmission rate, mode switch > 7 days, or initial programming mode besides DDD-CLS/DDD. Pacing characteristics were compared across DDD and DDD-CLS modes and between LBBAP and non-LBBAP lead locations.

RESULTS

A total of 5,395 patients were included. Baseline mean thoracic impedance was higher in non-LBBAP groups (P < 0.001), though changes over time were not significant (P = 0.336). Mean ventricular rate increased in the non-LBBAP DDD-CLS group but decreased in the DDD group (P < 0.001). Atrial arrhythmia burden was similar across pacing modes and lead locations, except for a higher burden in the non-LBBAP DDD-CLS group compared with non-LBBAP DDD.

CONCLUSIONS

In this large real-world cohort, CLS maintained stable physiological rate adaptation across pacing modes and lead locations. Minor differences in ventricular rate and thoracic impedance were observed, but overall CLS performance was consistent between LBBAP and non-LBBAP configurations, supporting its use in conduction system pacing.

Для цитирования: Yarmohammadi Hirad, Salehi Bobak, Shea Margaret, Harrell Camden, Miller Crystal, Hayes David. Efficacy of closed loop stimulation in left bundle branch area pacing.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2026. DOI: 10.1007/s10840-026-02430-7.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.