DOI 10.3390/diagnostics16162514

Медицинская библиотека · Diagnostics (Basel, Switzerland) · 2026

Comparison of LC-MS/MS and CLIA Methods for Vitamin D Measurement in an Unselected Outpatient Cohort: Statistical Evaluation and Impact on Clinical Classification.

Corbetta Carlo, Corsaro Luigi, Caradonna Eugenio, Rusconi Chiara, Rosa Ivana De, Diani Stefano, Gambaro Veniero, Grazia Ugo de, Maiocchi Alessandro, Ferrara Fulvio, Costantino Lucy, Roda Katia

СтатьяПеревод в очереди

Аннотация

BACKGROUND

This study aimed to compare a chemiluminescence immunoassay (CLIA; Siemens Atellica IM 1600) with an automated liquid chromatography-tandem mass spectrometry (LC-MS/MS) analyser (Thermo Scientific Cascadion SM) for serum 25-hydroxyvitamin D [25(OH)D] measurement in unselected outpatients, and its impact on clinical classification.

METHODS

25(OH)D was measured on 1064 paired serum samples (828 women, 236 men; age 3-111 years); both platforms participated in the Vitamin D Standardisation Program (VDSP). Agreement was assessed by the Wilcoxon signed-rank test, Passing-Bablok regression, Bland-Altman analysis on log-transformed values, Breusch-Pagan testing for heteroscedasticity, consistency-based intraclass correlation coefficient (ICC), Lin's concordance correlation coefficient, and Cohen's kappa with McNemar's test on six- and two-category clinical classifications.

RESULTS

Methods differed significantly (p<2.2×10-16). Passing-Bablok regression showed proportional bias (slope 1.33; 95% confidence interval [CI], 1.28-1.38; LC-MS/MS higher). Consistency ICC was 0.928 (95% CI, 0.919-0.936). Log-Bland-Altman bias was 35.5% (limits of agreement, -18.3% to +124.5%; heteroscedasticity, p=5.18×10-11). Cohen's kappa was 0.32 unweighted and 0.72 weighted on six clinical categories, and 0.56 on a two-class scheme (McNemar p<2.2×10-16). In discordant pairs, CLIA classified deficiency 118-fold more often than LC-MS/MS.

CONCLUSIONS

In our study, despite preserved rank ordering, a non-linear concentration-dependent bias and different limits of agreement caused systematic differences in the classification of vitamin D deficiency. Adoption of automated LC-MS/MS can substantially reclassify outpatients toward sufficiency, with implications for supplementation and laboratory harmonisation.

Для цитирования: Corbetta Carlo, Corsaro Luigi, Caradonna Eugenio, Rusconi Chiara, Rosa Ivana De, Diani Stefano, Gambaro Veniero, Grazia Ugo de, Maiocchi Alessandro, Ferrara Fulvio, Costantino Lucy, Roda Katia. Comparison of LC-MS/MS and CLIA Methods for Vitamin D Measurement in an Unselected Outpatient Cohort: Statistical Evaluation and Impact on Clinical Classification.. Diagnostics (Basel, Switzerland). 2026. DOI: 10.3390/diagnostics16162514.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.