DOI 10.1016/j.gofs.2018.11.012

Медицинская библиотека · Gynecologie, obstetrique, fertilite & senologie · 2018

[Correlation between automated optical reading and double human reading for two tests of fetal membranes rupture].

Four M, Chauffour C, Delabaere A, Pereira B, Sapin V, Gallot D

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

To assess concordance between double human and automated optical reading (AOR) concerning two biological tests for rupture of membranes (ROM)

METHODS

We conducted a monocentric, prospective, observational study comparing Actim Prom® (Alere SAS, Jouy-en Josas, France) and Hiprom Duo® (Fumouze, Levallois-Perret, France). Each test was performed simultaneously in patients with suspected ROM and read independently by 2 biologists and AOR device. ROM was clinically confirmed in case of recurrent leakage or spontaneous labour with no perceived membranes within 48hours.

RESULTS

Concerning Actim Prom®, concordance was 100 %, 92.5 % and 91.6 % between biologists, biologists-AOR device and biologists or AOR vs. clinical presentation respectively. Concerning Hiprom Duo®, concordance was 97.2 % between biologists, 97.2 % between biologist 1 and AOR, 95.3 % between biologist 2 and AOR, 63.5 % between clinical presentation and human reading, 62.3 % between clinical presentation and AOR. False positive cases were significantly associated with modified cervix (21 % vs. 46 %, P=0.006).

CONCLUSION

We demonstrated excellent correlation between biologists and good or excellent correlation between AOR and human reading supporting the use of AOR in clinical practice.

Для цитирования: Four M, Chauffour C, Delabaere A, Pereira B, Sapin V, Gallot D. [Correlation between automated optical reading and double human reading for two tests of fetal membranes rupture].. Gynecologie, obstetrique, fertilite & senologie. 2018. DOI: 10.1016/j.gofs.2018.11.012.