DOI 10.1080/14767058.2018.1519542

Медицинская библиотека · The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians · 2018

Comparison of reliability between uterocervical angle and cervical length measurements by various experienced operators using transvaginal ultrasound.

Pruksanusak Ninlapa, Sawaddisan Rapphon, Kor-Anantakul Ounjai, Suntharasaj Thitima, Suwanrath Chitkasaem, Geater Alan

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Purpose: To compare the reliability between uterocervical angle (UCA) and cervical length (CL) measurements by various experienced operators.

Methods

Transvaginal ultrasonographies (TVS) were performed in 102 pregnant women between 16°/7-24°/7 gestational weeks by different levels of experienced operators. For both CL and UCA measurements, intraobserver variability was assessed for each operator by examining the range between maximum and minimum measurements in each participant, compared to the mean of all three measurements. Interobserver variabilities were explored using Bland-Altman analysis. Intraclass correlation coefficients were used for both intraobserver and interobserver reliability.

Results

For intraobserver variability of the UCA, the ranges between maximum and minimum UCA measurements in operator 1 and 3 were 1.5º-34º and 2º-36º (n = 51), and in operators 2 and 3 were 0º-61º and 1º-25º (n = 51). Intraclass correlation coefficients (ICC) for intraobserver reliability were 0.90 for operator 1, 0.67 for operator 2 and 0.93 for operator 3. For interobserver variability of the UCA, the limits of agreement for mean UCA were -37.53º-38.41º and -36.27º-26.17º, and for maximum UCA were -39.47º-41.38º and -44.24º-22.9º in comparisons between operators 1 and 3, and operator 2 and 3, respectively. Intraclass correlation coefficients for mean UCA were 0.73 and 0.74, and for maximum UCA were 0.71 and 0.67 in comparisons between operators 1 and 3, and operator 2 and 3, respectively.

Conclusions

The UCA measurements had a higher intra- and interobserver reliabilities than the CL measurements.

Для цитирования: Pruksanusak Ninlapa, Sawaddisan Rapphon, Kor-Anantakul Ounjai, Suntharasaj Thitima, Suwanrath Chitkasaem, Geater Alan. Comparison of reliability between uterocervical angle and cervical length measurements by various experienced operators using transvaginal ultrasound.. The journal of maternal-fetal & neonatal medicine : the official journal of the European Association of Perinatal Medicine, the Federation of Asia and Oceania Perinatal Societies, the International Society of Perinatal Obstetricians. 2018. DOI: 10.1080/14767058.2018.1519542.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.