Brachytherapy · 2021

Quantifying clinical severity of physics errors in high-dose rate prostate brachytherapy using simulations.

СтатьяПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Nunez David Aramburu, Trager Michael, Beaudry Joel, Cohen Gilad N, Dauer Lawrence T, Gorovets Daniel, Hassan Rezaeian Nima, Kollmeier Marisa A, Leong Brian, McCann Patrick, Williamson Matthew, Zelefsky Michael J, Damato Antonio L

Аннотация

PURPOSE: To quantitatively evaluate through automated simulations the clinical significance of potential high-dose rate (HDR) prostate brachytherapy (HDRPB) physics errors selected from our internal failure-modes and effect analysis (FMEA).

METHODS AND MATERIALS: A list of failure modes was compiled and scored independently by 8 brachytherapy physicists on a one-to-ten scale for severity (S), occurrence (O), and detectability (D), with risk priority number (RPN) = SxOxD. Variability of RPNs across observers (standard deviation/average) was calculated. Six idealized HDRPB plans were generated, and error simulations were performed: single (N = 1722) and systematic (N = 126) catheter shifts (craniocaudal; -1cm:1 cm); single catheter digitization errors (tip and connector needle-tips displaced independently in random directions; 0.1 cm:0.5 cm; N = 44,318); and swaps (two catheters swapped during digitization or connection; N = 528). The deviations due to each error in prostate D90%, urethra D20%, and rectum D1cm3 were analyzed using two thresholds: 5-20% (possible clinical impact) and >20% (potentially reportable events).

RESULTS: Twenty-nine relevant failure modes were described. Overall, RPNs ranged from 6 to 108 (average ± 1 standard deviation, 46 ± 23), with responder variability ranging from 19% to 184% (average 75% ± 30%). Potentially reportable events were observed in the simulations for systematic shifts >0.4 cm for prostate and digitization errors >0.3 cm for the urethra and >0.4 cm for rectum. Possible clinical impact was observed for catheter swaps (all organs), systematic shifts >0.2 cm for prostate and >0.4 cm for rectum, and digitization errors >0.2 cm for prostate and >0.1 cm for urethra and rectum.

CONCLUSIONS: A high variability in RPN scores was observed. Systematic simulations can provide insight in the severity scoring of multiple failure modes, supplementing typical FMEA approaches.