Diagnostic Performance of Two Point-of-Care Tests for Anti-HCV Detection.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
da Rosa Lígia, Dantas-Corrêa Esther Buzaglo, Narciso-Schiavon Janaína Luz, Schiavon Leonardo de Lucca
Аннотация
BACKGROUND: Besides the great importance of the issue in terms of public health, there is a lack of studies evaluating the performance of several of the currently used point of care tests (POCTs) for the detection of anti-HCV.
OBJECTIVES: To investigate the performance of two POCTs for anti-HCV detection and to assess the impact of the reading time on diagnostic performance.
PATIENTS AND METHODS: A total of 307 subjects were divided into three groups (1- HCV infected; 2- other chronic liver diseases; and 3- controls). The POCTs HCV Rapid Test Bioeasy® and Imuno-Rapido HCV® were read at 3, 5, 10, 15, 20 and 30 minutes. The sensitivity and specificity of the POCTs were calculated in relation to anti-HCV detection by chemiluminescence.
RESULTS: Valid results were obtained for all tests. When compared to the chemiluminescence, both tests showed sensitivity of 97.1% and specificity of 100%. No changes in the sensitivity or specificity of the tests were observed at different reading times and when patients with other chronic liver diseases were evaluated as a control group.
CONCLUSIONS: The POCTs evaluated in this study showed high sensitivity and specificity, with no change in the performance after the third minute of reading.
OBJECTIVES: To investigate the performance of two POCTs for anti-HCV detection and to assess the impact of the reading time on diagnostic performance.
PATIENTS AND METHODS: A total of 307 subjects were divided into three groups (1- HCV infected; 2- other chronic liver diseases; and 3- controls). The POCTs HCV Rapid Test Bioeasy® and Imuno-Rapido HCV® were read at 3, 5, 10, 15, 20 and 30 minutes. The sensitivity and specificity of the POCTs were calculated in relation to anti-HCV detection by chemiluminescence.
RESULTS: Valid results were obtained for all tests. When compared to the chemiluminescence, both tests showed sensitivity of 97.1% and specificity of 100%. No changes in the sensitivity or specificity of the tests were observed at different reading times and when patients with other chronic liver diseases were evaluated as a control group.
CONCLUSIONS: The POCTs evaluated in this study showed high sensitivity and specificity, with no change in the performance after the third minute of reading.