DOI 10.1007/s00330-016-4340-4

Медицинская библиотека · European radiology · 2016

Three-dimensional C-arm CT-guided transjugular intrahepatic portosystemic shunt placement: Feasibility, technical success and procedural time.

Ketelsen Dominik, Groezinger Gerd, Maurer Michael, Lauer Ulrich M, Grosse Ulrich, Horger Marius, Nikolaou Konstantin, Syha Roland

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Establishment of transjugular intrahepatic portosystemic shunts (TIPS) constitutes a standard procedure in patients suffering from portal hypertension. The most difficult step in TIPS placement is blind puncture of the portal vein. This study aimed to evaluate three-dimensional mapping of portal vein branches and targeted puncture of the portal vein.

METHODS

Twelve consecutive patients suffering from refractory ascites by liver cirrhosis were included in this retrospective study to evaluate feasibility, technical success and procedural time of C-arm CT-targeted puncture of the portal vein. As a control, 22 patients receiving TIPS placement with fluoroscopy-guided blind puncture were included to compare procedural time.

RESULTS

Technical success could be obtained in 100 % of the study group (targeted puncture) and in 95.5 % of the control group (blind puncture). Appropriate, three-dimensional C-arm CT-guided mapping of the portal vein branches could be achieved in all patients. The median number of punctures in the C-arm CT-guided study group was 2 ± 1.3 punctures. Procedural time was significantly lower in the study group (14.8 ± 8.2 min) compared to the control group (32.6 ± 22.7 min) (p = 0.02).

CONCLUSIONS

C-arm CT-guided portal vein mapping is technically feasible and a promising tool for TIPS placement resulting in a significant reduction of procedural time.

KEY POINTS: • C-arm CT-mapping of the portal vein for 3D TIPS guidance is feasible. • Targeted punctures of the portal vein by C-arm CT reduce procedural time. • A decreased number of punctures could improve patient safety.

Для цитирования: Ketelsen Dominik, Groezinger Gerd, Maurer Michael, Lauer Ulrich M, Grosse Ulrich, Horger Marius, Nikolaou Konstantin, Syha Roland. Three-dimensional C-arm CT-guided transjugular intrahepatic portosystemic shunt placement: Feasibility, technical success and procedural time.. European radiology. 2016. DOI: 10.1007/s00330-016-4340-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.