Magnetic resonance cholangiopancreatography with GRASE sequence at 3.0T: does it improve image quality and acquisition time as compared with 3D TSE?
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yoshida Morikatsu, Nakaura Takeshi, Inoue Taihei, Tanoue Shota, Takada Sentaro, Utsunomiya Daisuke, Tsumagari Shota, Harada Kazunori, Yamashita Yasuyuki
Аннотация
OBJECTIVES: The current study evaluated the clinical usefulness of the gradient and spin-echo (GRASE) sequence with single breath-hold in 3.0 T magnetic resonance cholangiopancreatography (MRCP). We compared the acquisition time and image quality between GRASE and breath navigator-triggered 3D turbo spin echo (3D TSE).
METHODS: We examined 54 consecutive patients who underwent MRCP with GRASE and 3D TSE. We compared the image acquisition time and contrast-to-noise ratio (CNR) between the common bile duct (CBD) and liver. Overall image quality, blurring, motion artifacts and CBD visibility were scored on a 4-point scale by two radiologists. Paired t-tests were used to compare the variables.
RESULTS: The mean image acquisition time was 95 % shorter with the GRASE than with 3D TSE (GRASE: 20 s; 3D TSE: 6 min 27 s). The CNR of GRASE was significantly higher than that of 3D TSE (GRASE: 25.4 ± 13.9 vs. 3D TSE: 18.2 ± 9.6, p < 0.01). All qualitative scores for GRASE were significantly better than those for 3D TSE.
CONCLUSIONS: 3.0 T MRCP with GRASE sequence with single breath-hold significantly improved the CNR of CBD with a 95 % shorter acquisition time compared with conventional 3D MRCP with 3D TSE.
KEY POINTS: • MRCP acquisition time was 95% shorter with GRASE than with 3D TSE. • Overall image quality of GRASE was significantly better than 3D TSE. • Pancreaticobiliary tree visibility with GRASE was better than that with 3D TSE.
METHODS: We examined 54 consecutive patients who underwent MRCP with GRASE and 3D TSE. We compared the image acquisition time and contrast-to-noise ratio (CNR) between the common bile duct (CBD) and liver. Overall image quality, blurring, motion artifacts and CBD visibility were scored on a 4-point scale by two radiologists. Paired t-tests were used to compare the variables.
RESULTS: The mean image acquisition time was 95 % shorter with the GRASE than with 3D TSE (GRASE: 20 s; 3D TSE: 6 min 27 s). The CNR of GRASE was significantly higher than that of 3D TSE (GRASE: 25.4 ± 13.9 vs. 3D TSE: 18.2 ± 9.6, p < 0.01). All qualitative scores for GRASE were significantly better than those for 3D TSE.
CONCLUSIONS: 3.0 T MRCP with GRASE sequence with single breath-hold significantly improved the CNR of CBD with a 95 % shorter acquisition time compared with conventional 3D MRCP with 3D TSE.
KEY POINTS: • MRCP acquisition time was 95% shorter with GRASE than with 3D TSE. • Overall image quality of GRASE was significantly better than 3D TSE. • Pancreaticobiliary tree visibility with GRASE was better than that with 3D TSE.