Anterior chamber angle measurement with optical coherence tomography: intraobserver and interobserver variability.

Müller Maya, Dahmen Gerlinde, Pörksen Erk, Geerling Gerd, Laqua Horst, Ziegler Andreas, Hoerauf Hans

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess intraobserver and interobserver variability of anterior segment optical coherence tomography (AS-OCT) as an objective diagnostic tool to quantify the anterior chamber angle and opening width.

SETTING

University Eye Clinic, Lübeck, Germany.

METHODS

The anterior chamber angle and opening width were assessed in 18 eyes of 9 healthy volunteers by 2 observers. Intraobserver reproducibility was evaluated by calculating an intraclass correlation coefficient (ICC) in a mixed model. Each observer had a separate model using information from 5 scans. Interobserver variability was determined by Bland-Altman analysis. The ICC was calculated in a mixed model using a residual maximum likelihood method. The results of 3 and 5 repeated scans were evaluated to indicate the change to 1 measurement application.

RESULTS

The mean anterior chamber angle measurement was 35.9 degrees +/- 5.7 (SD) for observer A and 36.2 +/- 5.7 degrees for observer B. The ICC was 0.94 and 0.91, respectively. The mean opening width was 315 +/- 62 microm for observer A and 317 +/- 60 microm for observer B. The ICC was 0.97 and 0.93, respectively. Interobserver comparisons showed a mean difference between anterior chamber angle measurements of -0.27 +/- 1.6 degrees, a limit of agreement (LOA) interval from -3.52 to 2.98 degrees, and an ICC estimate of 0.96. The mean difference in opening width measurements was 2.40 +/- 12.40 microm, the LOA from -27.20 to 22.40 microm, and the estimated ICC 0.96. Using 1 instead of the mean of 5 measurements, the LOA range increased by 3.46 degrees for the anterior chamber angle and 30.0 microm for the opening width.

CONCLUSION

Anterior chamber angle and opening width measurements by OCT showed low intraobserver and interobserver variability, indicating OCT is a valuable technique for quantitative assessment that provides reproducible measurements and objective documentation by different examiners.

Для цитирования: Müller Maya, Dahmen Gerlinde, Pörksen Erk, Geerling Gerd, Laqua Horst, Ziegler Andreas, Hoerauf Hans. Anterior chamber angle measurement with optical coherence tomography: intraobserver and interobserver variability.. Journal of cataract and refractive surgery. 2006. DOI: 10.1016/j.jcrs.2006.07.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.