DOI 10.1097/mao.0b013e3181fcf167
Intraoperative monitoring of active middle ear implant function in patients with normal and pathologic middle ears.
Аннотация
OBJECTIVE
To describe a technique enabling the intraoperative assessment of the mechanical coupling of active middle ear implants in patients with normal or dysfunctional middle ears.
PATIENTS: Patients with sensorineural or mixed hearing loss subjected to revision surgery of an active middle ear implant.
INTERVENTION: Recording of compound action potentials (CAPs) of the auditory nerve in response to stimulation via an active middle ear implant during revision surgery.
MAIN OUTCOME MEASURES: CAP thresholds as a measure of mechanical coupling and system integrity.
CONCLUSION
Determining CAP thresholds may be suitable for identifying a proper mechanical coupling of active middle ear implants in patients with a dysfunctional middle ear, even if the ossicular chain is disrupted.