Moscarella Elisabetta, Ielasi Alfonso, Kim Won-Keun, Caminiti Rodolfo, Barbanti Marco, Giordano Arturo, Holzamer Andreas, Testa Luca, Adamo Marianna, Eitan Amnon, Estévez-Loureiro Rodrigo, Farkic Mihajlo, Stoyanov Nikolay, Leick Juergen, Kovacevic Mila, Tigges Eike, Mangieri Antonio, Scotti Andrea, Latib Azeem

СтатьяМногоцентровое исследованиеСравнительное исследованиеНаблюдательное исследованиеПеревод в очереди

Аннотация

BACKGROUND

With transcatheter aortic valve implantation (TAVI) now extending to lower-risk and younger patients, optimizing procedural and hemodynamic outcomes is critical. The Myval Octacor, a new balloon-expandable valve (BEV), was developed to improve outcomes by reducing paravalvular regurgitation (PVL), minimizing pacemaker implantation (PPI) rates, and enhancing hemodynamic performance. However, limited data are available comparing Myval Octacor to contemporary self-expanding supra-annular valves (SEVs) Evolut PRO/PRO+ and Acurate Neo2.

AIMS: This study aimed to compare the safety, efficacy, and short-term clinical outcomes of Myval Octacor with SEVs in transfemoral TAVI patients.

METHODS

We conducted a prospective, multicenter registry including patients treated with Myval Octacor, then compared them to SEV recipients from the NEOPRO2 registry. Propensity score matching adjusted for baseline differences between groups. The primary endpoint was 30-day Valve Academic Research Consortium-3 (VARC-3) device success. Secondary endpoints included technical success, valve performance, and early safety outcomes.

RESULTS

Among 252 Myval Octacor patients and 2175 SEV patients, 90 matched pairs were compared. Myval Octacor patients showed higher 30-day VARC-3 device success than SEVs patients (97% vs. 88%, p = 0.024), primarily due to a numerically lower rate of moderate-to-severe PVL (1% vs. 7%, p = 0.06). The Octacor group also exhibited larger indexed effective orifice areas. Mortality, stroke, PPI, and myocardial infarction rates were similar between groups at 30 days.

CONCLUSIONS

The Myval Octacor demonstrated comparable early safety and efficacy to supra-annular SEVs, with advantages in device success rate and hemodynamic performance. Adequately sized randomized study is required to confirm these findings.

Для цитирования: Moscarella Elisabetta, Ielasi Alfonso, Kim Won-Keun, Caminiti Rodolfo, Barbanti Marco, Giordano Arturo, Holzamer Andreas, Testa Luca, Adamo Marianna, Eitan Amnon, Estévez-Loureiro Rodrigo, Farkic Mihajlo, Stoyanov Nikolay, Leick Juergen, Kovacevic Mila, Tigges Eike, Mangieri Antonio, Scotti Andrea, Latib Azeem. Balloon-Expandable Myval Octacor Versus Self-Expanding Evolut PRO/PRO+ and Acurate Neo2: Short-Term Outcomes From Propensity-Matched Analysis.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.70097.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.