European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2021

Speech perception and hearing effort using a new active middle ear implant audio processor.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Rahne Torsten, Fröhlich Laura, Wagner Luise, Kropp Miriam Hannah, Müller Alexander

Аннотация

PURPOSE: The Vibrant Soundbridge (VSB) was introduced in 1996, and the fourth generation of the audio processor recently released. This clinical study evaluates the audiological performance and subjective satisfaction of the new SAMBA 2 audio processor compared to its predecessor, SAMBA.

METHOD: Fifteen VSB users tested both audio processors for approximately 3 weeks. Air conduction and bone conduction thresholds and unaided and aided sound field thresholds were measured with both devices. Speech performance in quiet (Freiburg monosyllables) and noise (OLSA) was evaluated as well as subjective listening effort (ACALES) and questionnaire outcomes (SSQ12 and APSQ). In addition, data from 16 subjects with normal hearing were gathered on sound field tests and ACALES.

RESULTS: Both audio processors showed substantial improvement compared to the unaided condition. The SAMBA and SAMBA 2 had comparable performance in sound filed thresholds, while the SAMBA 2 was significantly better in speech in quiet, speech in noise, reduced listening effort, and improved subjective satisfaction compared with the SAMBA.

CONCLUSION: The SAMBA 2 audio processor, compared to its predecessor SAMBA, offers improved performance throughout the parameters investigated in this study. Patients with a VSB implant would benefit from an upgrade to SAMBA 2.