Variables decreasing tip movement of peripherally inserted central catheters in pediatric patients.

Gnannt Ralph, Connolly Bairbre L, Parra Dimitri A, Amaral Joao, Moineddin Rahim, Thakor Avnesh S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The position of the tip of a peripherally inserted central catheter (PICC) is crucial; malposition can lead to malfunction of the line or life-threatening events (e.g., arrhythmias, perforation).

OBJECTIVE

To determine what factors other than arm position and accessed vein might influence the tip position of a PICC.

MATERIALS AND

METHODS

Inclusion criteria were upper limb PICC placement, body weight <20 kg, intraoperative imaging with the arm in 0°, 45° and 90° abduction and an arm view marking the skin entry site relative to the shoulder. Evaluated variables included patient demographics, and PICC and insertion site characteristics. We measured central tip movement in rib units.

RESULTS

We included 112 children who received a PICC (42 girls/70 boys, mean age 31±13 months, mean weight 6.5±4.9 kg). The overall range of central tip movement was -1 to +4 rib units (mean +0.8±0.7 rib units). Silicone PICCs moved significantly less than polyurethane PICCs (P<0.05). PICCs placed in the cephalic vein moved significantly less than those placed in other veins (P<0.05). Patient demographics and PICC characteristics (size, number of lumens, left or right arm accessed, length of the line) did not influence the range of central tip movement of a PICC (P>0.05).

CONCLUSION

Silicone PICCs and PICCs inserted into the cephalic vein move less than PICCs made of polyurethane and PICCs inserted into the brachial and basilic veins. These findings might assist operators in deciding which PICC to place in children in a given clinical context.

Для цитирования: Gnannt Ralph, Connolly Bairbre L, Parra Dimitri A, Amaral Joao, Moineddin Rahim, Thakor Avnesh S. Variables decreasing tip movement of peripherally inserted central catheters in pediatric patients.. Pediatric radiology. 2016. DOI: 10.1007/s00247-016-3648-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.