Improved Assessment of Bolus Clearance in Patients With Achalasia Using High-Resolution Impedance Manometry.

Carlson Dustin A, Beveridge Claire A, Lin Zhiyue, Balla Michelle, Gregory Dyanna, Tye Michael, Ritter Katherine, Kahrilas Peter J, Pandolfino John E

Оценочное исследованиеСтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

BACKGROUND & AIMS: Esophageal retention is typically evaluated by timed-barium esophagram in patients treated for achalasia. Esophageal bolus clearance can also be evaluated using high-resolution impedance manometry. We evaluated the associations of conventional and novel high-resolution impedance manometry metrics, esophagram, and patient-reported outcomes (PROs) in achalasia.

METHODS

We performed a prospective study of 70 patients with achalasia (age, 20-81 y; 30 women) treated by pneumatic dilation or myotomy who underwent follow-up evaluations from April 2013 through December 2015 (median, 12 mo after treatment; range, 3-183 mo). Patients were assessed using timed-barium esophagrams, high-resolution impedance manometry, and PROs, determined from Eckardt scores (the primary outcome) and the brief esophageal dysphagia questionnaire. Barium column height was measured from esophagrams taken 5 minutes after ingestion of barium (200 mL). Impedance-manometry was analyzed for bolus transit (dichotomized) and with a customized MATLAB program (The MathWorks, Inc, Natick, MA) to calculate the esophageal impedance integral (EII) ratio.

RESULTS

Optimal cut points to identify a good PRO (defined as Eckardt score of ≤3) were esophagram barium column height of 3 cm (identified patients with a good PRO with 63% sensitivity and 75% specificity) and an EII ratio of 0.41 (identified patients with a good PRO with 83% sensitivity and 75% specificity). Complete bolus transit identified patients with a good PRO with 28% sensitivity and 75% specificity. Of the 25 patients who met these cut points for both esophagram barium column height and EII ratio, 23 (92%) had a good PRO. Of the 17 patients who met neither cut point, 14 (82%) had a poor PRO (Eckardt score above 3).

CONCLUSIONS

In a prospective study of 70 patients with achalasia, we found EII ratio identified patients with good PROs with higher levels of sensitivity (same specificity) than timed-barium esophagram or impedance-manometry bolus transit assessments. The EII ratio should be added to achalasia outcome evaluations that involve high-resolution impedance manometry as an independent measure and to complement timed-barium esophagram.

Для цитирования: Carlson Dustin A, Beveridge Claire A, Lin Zhiyue, Balla Michelle, Gregory Dyanna, Tye Michael, Ritter Katherine, Kahrilas Peter J, Pandolfino John E. Improved Assessment of Bolus Clearance in Patients With Achalasia Using High-Resolution Impedance Manometry.. Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association. 2018. DOI: 10.1016/j.cgh.2017.11.019.