BACKGROUND AND AIM: In patients with irretrievable or intractable bile duct stone, temporary insertion of a plastic stent (PS) followed by further endoscopic retrograde cholangiopancreatography (ERCP) or surgery has been recommended as a 'bridge' therapy. However, the exact mechanism of stone fragmentation has not been discovered. The aim of the present study was to evaluate whether PS shape can facilitate stone fragmentation.

METHODS: Using a new in vitro bile flow phantom model, we compared the friction effect among three different PS groups (straight PS group, double pigtail-shaped PS group, and screw-shaped PS group) and a control group. Each group had 10 silicon tube blocks that separately contained one stone and two PS. The control group had 10 blocks each with only a stone and no PS. We carried out analysis of the friction effect by stone weight and volume changes among the groups, excluding fragmented stones.

RESULTS: After 8 weeks, complete fragmentation was noted in one out of 34 cholesterol stones (2.9%) and in four out of six pigmented stones (66.7%). Fragmentation tended to be more prominent in the screw-shaped PS group than in the straight PS group, double pigtail-shaped group, and control group (volume change: -11.33%, 7.94%, 4.43%, and 2.05%, respectively, P = 0.1390; weight change: -9.30%, 0.71%, -0.10%, and -1.23%, respectively, P = 0.3553).

CONCLUSION: Stone fragmentation may be induced by PS friction effect. Also, screw-shaped plastic stents may improve friction effect. These results may help guide future PS development and clinical decisions.