Fully implantable Otologics MET Carina(™) device for the treatment of sensorineural and mixed hearing loss: Audio-otological results.
Аннотация
CONCLUSION
Our results attest that the Otologics MET Carina(™) is a viable treatment for moderate to severe sensorineural hearing loss (SNHL) and for cases of mixed hearing loss, and that in selected cases it could represent an alternative to conventional hearing aids.
OBJECTIVES
To describe our experience with the fully implantable Carina(™) in eight adult patients, seven with moderate to severe SNHL and one with mixed hearing loss.
METHODS
Eight implanted adult patients were submitted to a comprehensive audiological evaluation.
RESULTS
We did not record any surgical relevant complication in any of the patients, nor any significant postoperative variation in hearing thresholds, for air conduction or bone conduction, indicating the absence of surgical damage to the cochlea. All the patients demonstrated improvements in speech perception abilities with the device functioning and reported subjective benefits. With regard to the postoperative adverse effects, we had problems with feedback noise, which resolved with minor fitting adjustments in seven cases, while it required a second surgery to change the microphone position in the other patient. In one case a minimal extrusion of the microphone cable occurred requiring a revision surgery, a device failure occurred in one case, requiring substitution, and one patient decided on explantation of the device owing to psychological problems.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.