MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/jmri.22066

Медицинская библиотека · Journal of magnetic resonance imaging : JMRI · 2010

Repeatability of magnetic resonance elastography for quantification of hepatic stiffness.

Hines Catherine D G, Bley Thorsten A, Lindstrom Mary J, Reeder Scott B

Оценочное исследованиеСтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the sources of variability of MRE hepatic stiffness measurements using healthy volunteers and patients and to calculate the minimum change required for statistical significance. Hepatic stiffness measured with magnetic resonance elastography (MRE) has demonstrated tremendous potential as a noninvasive surrogate of hepatic fibrosis, although the underlying repeatability of MRE for longitudinal tracking of liver disease has not been documented.

MATERIALS AND

METHODS

MRE stiffness measurements from 20 healthy volunteers and 10 patients were obtained twice on the same day, and repeated 2-4 weeks later for volunteers in this institutional review board-approved study. A linear mixed effects model was used to estimate the component sources of variability in the data.

RESULTS

The standard deviation of MRE measurements of the same individual on different days is 11.9% (percent of the measured stiffness) using the same reader and 12.0% using different readers. The standard deviation of the difference between two measurements (i.e., longitudinal change in an individual) is 17.4%; the corresponding 95% confidence interval for zero change is (-27.0%, 37.0%).

CONCLUSION

MRE is a repeatable method for quantifying liver stiffness. Using the described MRE technique, changes greater than 37.0% of the smaller measured stiffness value represent meaningful changes in longitudinal liver stiffness measurements.

Для цитирования: Hines Catherine D G, Bley Thorsten A, Lindstrom Mary J, Reeder Scott B. Repeatability of magnetic resonance elastography for quantification of hepatic stiffness.. Journal of magnetic resonance imaging : JMRI. 2010. DOI: 10.1002/jmri.22066.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.