MedElevator Войти в исследования
← Поиск публикаций

DOI 10.12455/j.issn.1671-7104.240480

Медицинская библиотека · Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation · 2025

[Analysis of the Influence of Different Scanning Conditions of Medical Linear Accelerator CBCT on Image Quality].

Liu Li, Ye Chengwei, Yuan Jianjun, Luo Yingui, Luo Zhiyao, Zeng Wei, Li Ling, Liu Huan, Liu Yan

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To investigate the influence of different scanning conditions on the image quality of medical electron accelerator cone-beam computed tomography (CBCT) and provide a reference for the selection of scanning conditions for different body parts. Methods Set different scanning conditions, the Catphan 503 phantom was scanned using CBCT parameters to analyze the influence of spatial resolution, noise, uniformity, spatial geometric accuracy, and low-contrast resolution on the image quality of CBCT.

RESULTS

For the head, chest, and abdomen, with the increase in scanning parameter values, the noise value decreased by 47.4%, 26.1%, and 51.3% respectively, and the uniformity values decreased by 30.2%, 26.6%, and 47.9% respectively. The low-contrast resolution values decreased by 50.6%, 34.2%, and 12.0%. The influence of different scanning conditions on spatial geometric accuracy and spatial resolution is not significant.

CONCLUSION

Different scanning parameters have a certain influence on the image quality of medical electron accelerator CBCT. Lower scanning parameters can be selected based on individual patients to reduce the additional radiation dose, providing a reference for the safe application of CBCT image guidance in radiotherapy.

Для цитирования: Liu Li, Ye Chengwei, Yuan Jianjun, Luo Yingui, Luo Zhiyao, Zeng Wei, Li Ling, Liu Huan, Liu Yan. [Analysis of the Influence of Different Scanning Conditions of Medical Linear Accelerator CBCT on Image Quality].. Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation. 2025. DOI: 10.12455/j.issn.1671-7104.240480.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.