Clinical Application of Dual-Energy Spectral Computed Tomography in Detecting Cholesterol Gallstones From Surrounding Bile.

Yang Chuang-Bo, Zhang Shuang, Jia Yong-Jun, Duan Hai-Feng, Ma Guang-Ming, Zhang Xi-Rong, Yu Yong, He Tai-Ping

СтатьяПеревод пока не подготовлен

Аннотация

RATIONALE AND

OBJECTIVE

This study aimed to investigate the clinical value of spectral computed tomography (CT) in the detection of cholesterol gallstones from surrounding bile.

MATERIALS AND

METHODS

This study was approved by the institutional review board. The unenhanced spectral CT data of 24 patients who had surgically confirmed cholesterol gallstones were analyzed. Lipid concentrations and CT numbers were measured from fat-based material decomposition image and virtual monochromatic image sets (40-140 keV), respectively. The difference in lipid concentration and CT number between cholesterol gallstones and the surrounding bile were statistically analyzed. Receiver operating characteristic analysis was applied to determine the diagnostic accuracy of using lipid concentration to differentiate cholesterol gallstones from bile.

RESULTS

Cholesterol gallstones were bright on fat-based material decomposition images yielding a 92% detection rate (22 of 24). The lipid concentrations (552.65 ± 262.36 mg/mL), CT number at 40 keV (-31.57 ± 16.88 HU) and 140 keV (24.30 ± 5.85 HU) for the cholesterol gallstones were significantly different from those of bile (-13.94 ± 105.12 mg/mL, 12.99 ± 9.39 HU and 6.19 ± 4.97 HU, respectively). Using 182.59 mg/mL as the threshold value for lipid concentration, one could obtain sensitivity of 95.5% and specificity of 100% with accuracy of 0.994 for differentiating cholesterol gallstones from bile.

CONCLUSIONS

Virtual monochromatic spectral CT images at 40 keV and 140 keV provide significant CT number differences between cholesterol gallstones and the surrounding bile. Spectral CT provides an excellent detection rate for cholesterol gallstones.

Для цитирования: Yang Chuang-Bo, Zhang Shuang, Jia Yong-Jun, Duan Hai-Feng, Ma Guang-Ming, Zhang Xi-Rong, Yu Yong, He Tai-Ping. Clinical Application of Dual-Energy Spectral Computed Tomography in Detecting Cholesterol Gallstones From Surrounding Bile.. Academic radiology. 2016. DOI: 10.1016/j.acra.2016.10.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.