Evaluation on Effectiveness of a New System as well as Analysis on Optimal Horizontal Eye Position for Vertical Video Head Impulse Test.
Аннотация
OBJECTIVE
To compare the performances among three different systems for video head impulse test (vHIT), and to identify an optimal target angle for precisely evaluating the function of vertical semicircular canals in vHIT.
METHODS
A two-center prospective study was done. Participants were sit 1.2 m away from the wall in a noise-proved room that dedicated for vHIT experiments. During the comparison experiments, similar settings were ensured in both hospitals, with the same distance to wall and angle of staring. For each equipment, the procedures followed the developers' recommendations. The same examiner performed the comparison between two systems in one location. For the eye-position projects, targets were placed on the wall sequentially at the pre-marked lines for different angles. For the comparison projects, 9 and 13 participants were recruited, respectively. Any participant with otologic or vestibular disorders was excluded. A total of 26 healthy participants were recruited in the eye-position experiments, 16 of which were further involved in inter-examiner tests.
RESULTS
Our evaluations of three different systems showed that a new vHIT system, VertiGoggles® ZT-VNG-I (VG) performed as good as the long-tested Otometrics® ICS impulse (Oto) and EyeSeeCam® (ESC). During the comparison, we validated 25-degree, instead of right ahead at 0 degree, is a better place to set the targets when torsion was applied at vertical semicircular canal planes.
CONCLUSION
The new VG system is good for clinical practices. Furthermore, we proposed a new protocol to set the targets 25 degrees from right ahead after tilting head 45 degrees to evaluate vertical canals during vHIT.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.