New Tool for Monitoring Molecular Response in Patients With Chronic Myeloid Leukemia.

Badar Talha, Luthra Rajyalakshmi, Kantarjian Hagop, Jabbour Elias, Borthakur Gautam, Garcia-Manero Guillermo, Huang Xuelin, Singh Rajesh, Alvarez Brittany, Austermiller Bradley, Morrison Tom B, Patel Keyur P, Cortes Jorge

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Chronic myeloid leukemia treatment monitoring using polymerase chain reaction-based peripheral blood testing of t9;22 BCR-ABL1 provides improved test sensitivity over cytology but suffers from inadequate standardization in most laboratories due to variations inherent in the existing polymerase chain reaction methodologies. We performed the initial analytic performance evaluation of a novel competitive template-based peripheral blood b2a2/b3a2 transcript abundance method, called standardized nucleic acid quantification (SNAQ) test, with hypothesis that this will produced more consistent results with less frequent interlaboratory variations.

MATERIALS AND

METHODS

Thirty-six chronic myeloid leukemia patients treated at our institution were enrolled. We compared SNAQ test with 2 laboratory developed test at the MD Anderson molecular diagnostic laboratory and Cancer Genetics Institute for analyzing BCR-ABL1 from peripheral blood samples. Each test result (n=36) was ranked against all the other samples tested by the same method.

RESULTS

The Pearson correlation between SNAQ and laboratory developed test done at 2 labs was met by correlations of 0.97, 0.96, 0.96, and 0.94. Analysis of variance of log %BCR-ABL1 interlaboratory results indicated no significant difference (P=0.98). Post hoc analysis of method agreement showed the SNAQ method had a 95% limit of agreement of ±3-fold between laboratories.

CONCLUSIONS

In this pilot study, SNAQ methodology performed consistent with half-log accuracy. Additional studies from a larger sample size and correlation with clinical outcomes are required to confirm this observation.

Для цитирования: Badar Talha, Luthra Rajyalakshmi, Kantarjian Hagop, Jabbour Elias, Borthakur Gautam, Garcia-Manero Guillermo, Huang Xuelin, Singh Rajesh, Alvarez Brittany, Austermiller Bradley, Morrison Tom B, Patel Keyur P, Cortes Jorge. New Tool for Monitoring Molecular Response in Patients With Chronic Myeloid Leukemia.. Applied immunohistochemistry & molecular morphology : AIMM. 2019. DOI: 10.1097/pai.0000000000000526.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.