Performing Quality Control on Magnetic Resonance Imaging Liver Fat/Iron Quantification Studies: A Critical Requirement.

Hajibonabi Farid, Sharma Puneet, Davarpanah Amir H, Balthazar Patricia, Moreno Courtney C, Pectasides Melina, Nandwana Sadhna B

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Nonalcoholic fatty liver and iron overload can lead to cirrhosis requiring early detection. Magnetic resonance (MR) imaging utilizing chemical shift-encoded sequences and multi-Time of Echo single-voxel spectroscopy (SVS) are frequently used for assessment. The purpose of this study was to assess various quality factors of technical acceptability and any deficiencies in technologist performance in these fat/iron MR quantification studies.

METHODS

Institutional review board waived retrospective quality improvement review of 87 fat/iron MR studies performed over a 6-month period was evaluated. Technical acceptability/unacceptability for chemical shift-encoded sequences (q-Dixon and IDEAL-IQ) included data handling errors (missing maps), liver field coverage, fat/water swap, motion, or other artifacts. Similarly, data handling (missing table/spectroscopy), curve-fit, fat- and water-peak separation, and water-peak sharpness were evaluated for SVS technical acceptability.

RESULTS

Data handling errors were found in 11% (10/87) of studies with missing maps or entire sequence (SVS or q-Dixon). Twenty-seven percent (23/86) of the q-Dixon/IDEAL-IQ were technically unacceptable (incomplete liver-field [39%], other artifacts [35%], significant/severe motion [18%], global fat/water swap [4%], and multiple reasons [4%]). Twenty-eight percent (21/75) of SVS sequences were unacceptable (water-peak broadness [67%], poor curve-fit [19%] overlapping fat and water peaks [5%], and multiple reasons [9%]).

CONCLUSIONS

A high rate of preventable errors in fat/iron MR quantification studies indicates the need for routine quality control and evaluation of technologist performance and technical deficiencies that may exist within a radiology practice. Potential solutions such as instituting a checklist for technologists during each acquisition procedure and routine auditing may be required.

Для цитирования: Hajibonabi Farid, Sharma Puneet, Davarpanah Amir H, Balthazar Patricia, Moreno Courtney C, Pectasides Melina, Nandwana Sadhna B. Performing Quality Control on Magnetic Resonance Imaging Liver Fat/Iron Quantification Studies: A Critical Requirement.. Journal of computer assisted tomography. 2023. DOI: 10.1097/rct.0000000000001471.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.