MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/01945998221116815

Медицинская библиотека · Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery · 2023

Does Prophylactic Replacement of Voice Prosthesis Make Sense? A Study to Predict Prosthesis Lifetime.

Heirman Anne N, van der Noort Vincent, van Son Rob, Petersen Japke F, van der Molen Lisette, Halmos Gyorgy B, Dirven Richard, van den Brekel Michiel W M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Voice prosthesis leakage significantly affects the quality of life of patients undergoing laryngectomy, causing insecurity and frequent unplanned hospital visits and costs. In this study, the concept of prophylactic voice prosthesis replacement was explored to prevent leakages.

STUDY

DESIGN

Retrospective cohort study.

SETTING

Tertiary hospital.

METHODS

This study included all patients who underwent laryngectomy between 2000 and 2012 in the Netherlands Cancer Institute. Device lifetimes and voice prosthesis replacements of a retrospective cohort were used to calculate the number of needed voice prostheses per patient per year to prevent 70% of the leakages by prophylactic replacement. Various strategies for the timing of prophylactic replacement were considered: adaptive strategies based on the individual patient's history of replacement and fixed strategies based on the results of patients with similar voice prosthesis or treatment characteristics.

RESULTS

Patients used a median 3.4 voice prostheses per year (range, 0.1-48.1). We found high inter- and intrapatient variability in device lifetime. When prophylactic replacement is applied, this would become a median 9.4 voice prostheses per year, which means replacement every 38 days, implying >6 additional voice prostheses per patient per year. The individual adaptive model showed that preventing 70% of the leakages was impossible for most patients and only a median 25% can be prevented. Monte-Carlo simulations showed that prophylactic replacement is not feasible due to the high coefficient of variation (SD/mean) in device lifetime.

CONCLUSION

Based on our simulations, prophylactic replacement of voice prostheses is not feasible due to high inter- and intrapatient variation in device lifetime.

Для цитирования: Heirman Anne N, van der Noort Vincent, van Son Rob, Petersen Japke F, van der Molen Lisette, Halmos Gyorgy B, Dirven Richard, van den Brekel Michiel W M. Does Prophylactic Replacement of Voice Prosthesis Make Sense? A Study to Predict Prosthesis Lifetime.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2023. DOI: 10.1177/01945998221116815.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.