Local Delivery of Mometasone Furoate from an Eluting Endotracheal Tube Reduces Airway Morbidity Following Long-Term Animal Intubation.

Jahshan Forsan, Abu Ammar Aiman, Ertracht Offir, Eisenbach Netanel, Daoud Amani, Sela Eyal, Atar Shaul, Zussman Eyal, Fichtman Boris, Harel Amnon, Gruber Maayan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

upper airway complications are common sequelae of endotracheal tube (ETT) intubation, and systemic corticosteroids are considered a mainstay treatment for this problem. Drug-eluting ETT may present an attractive option for topical steroid delivery while avoiding systemic side effects and improving the therapeutic outcome. The objective of the present study is to evaluate the reduction of tube-related tracheal morbidity via a self-designed steroid-eluting ETT with controlled sustained release properties in an animal model.

METHODS

steroid-eluting ETTs were coated by poly(lactic-co-glycolic acid) -electrospun nanofibers loaded with mometasone furoate (MF) as a model drug. Animals were randomly assigned into three equal groups: non-intubated, blank-ETT, and loaded-ETT. The intubation interval was 1 week. Specimens were analyzed by histology, specific fibrosis staining, and scanning electron microscopy (SEM).

RESULTS

the blank-ETT group exhibited a significant increase in tracheal mucosal thickness compared to the loaded-ETT and control groups. Average tracheal mucosal thickness was 112 ± 34, 242 ± 49, and 113 ± 43 μm in the control, blank-ETT, and loaded-ETT groups, respectively. The blank-ETT group exhibited a significant increase in tracheal fibrosis compared to the loaded-ETT and control groups. Relative fibrosis values were 0.07 ± 0.05, 0.154 ± 0.1, and 0.0984 ± 0.084% for the control, blank-ETT, and loaded-ETT groups, respectively. While SEM imaging showed normal surface structures in the control group, intubated blank-ETT rats showed severe surface structural damage, whereas only mild damage was observed in the loaded-ETT group.

CONCLUSIONS

local sustained release of MF via a self-designed drug-eluting ETT is a potential therapeutic approach which may significantly reduce tube-related upper airway morbidity.

Для цитирования: Jahshan Forsan, Abu Ammar Aiman, Ertracht Offir, Eisenbach Netanel, Daoud Amani, Sela Eyal, Atar Shaul, Zussman Eyal, Fichtman Boris, Harel Amnon, Gruber Maayan. Local Delivery of Mometasone Furoate from an Eluting Endotracheal Tube Reduces Airway Morbidity Following Long-Term Animal Intubation.. ACS applied bio materials. 2021. DOI: 10.1021/acsabm.0c01526.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.