m+Medelevator.RU
← Поиск публикаций
Trends in hearing · 2025

A Prospective, Multicentre Case-Control Trial Examining Factors That Explain Variable Clinical Performance in Post Lingual Adult CI Recipients.

СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Dawson Pam, Fullerton Amanda, Krishnamoorthi Harish, Plant Kerrie, Cowan Robert, Buczak Nadine, Long Christopher, James Chris J, Sismono Fergio, Büchner Andreas

Аннотация

This study investigated which of a range of factors could explain performance in two distinct groups of experienced, adult cochlear implant recipients differentiated by performance on words in quiet: 72 with poorer word scores versus 77 with better word scores. Tests measured the potential contribution of sound processor mapping, electrode placement, neural health, impedance, cognitive, and patient-related factors in predicting performance. A systematically measured sound processor MAP was compared to the subject's walk-in MAP. Electrode placement included modiolar distance, basal and apical insertion angle, and presence of scalar translocation. Neural health measurements included bipolar thresholds, polarity effect using asymmetrical pulses, and evoked compound action potential (ECAP) measures such as the interphase gap (IPG) effect, total refractory time, and panoramic ECAP. Impedance measurements included trans impedance matrix and four-point impedance. Cognitive tests comprised vocabulary ability, the Stroop test, and the Symbol Digits Modality Test. Performance was measured with words in quiet and sentence in noise tests and basic auditory sensitivity measures including phoneme discrimination in noise and quiet, amplitude modulation detection thresholds and quick spectral modulation detection. A range of predictor variables accounted for between 33% and 60% of the variability in performance outcomes. Multivariable regression analyses showed four key factors that were consistently predictive of poorer performance across several outcomes: substantially underfitted sound processor MAP thresholds, higher average bipolar thresholds, greater total refractory time, and greater IPG offset. Scalar translocation, cognitive variables, and other patient related factors were also significant predictors across more than one performance outcome.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты