Panbio antigen rapid test is reliable to diagnose SARS-CoV-2 infection in the first 7 days after the onset of symptoms.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Linares Manuel, Pérez-Tanoira Ramón, Carrero Ana, Romanyk Juan, Pérez-García Felipe, Gómez-Herruz Peña, Arroyo Teresa, Cuadros Juan
Аннотация
BACKGROUND: RT-qPCR is the current recommended laboratory method to diagnose SARS-CoV-2 acute infection, several factors such as requirement of special equipment, time consuming, high cost and skilled staff limit the use of these techniques. A more rapid and high-throughput method is essential.
METHODS: We analyzed clinical data and nasopharyngeal samples, collected during September 2020, from patients attended at the emergency department of a secondary hospital and in two primary healthcare centers in Madrid. The performance of the Panbio™ COVID-19 AG Rapid Test Device for the detection of SARS-CoV-2 antigen was compared to RT-qPCR.
RESULTS: 255 nasopharyngeal swabs, including 150 from the emergency department and 105 from primary helthcare centers, were tested. 184 patients were symptomatic (72.1 %). Amongst the 60 positive RT-qPCR samples, 40 were detected by the rapid antigen test, given an overall sensitivity of 73.3 %. All the samples detected positive with the rapid antigen test were also positive with RT-qPCR. The median cycle threshold was 23.28 (IQR 18.5-30.16). Patients with less than seven days onset of symptoms showed a higher viral load, and sensitivity for rapid antigen test (86.5 %), compared to those with more days (sensitivity of 53.8 %)(p < 0.004).
CONCLUSIONS: The rapid antigen test evaluated in this study showed a high sensitivity and specificity in samples obtained during the first week of symptoms and with high viral loads. This assay seems to be an effective strategy for controlling the COVID-19 pandemic for the rapid identification and isolation of SARS-CoV-2 infected patients.
METHODS: We analyzed clinical data and nasopharyngeal samples, collected during September 2020, from patients attended at the emergency department of a secondary hospital and in two primary healthcare centers in Madrid. The performance of the Panbio™ COVID-19 AG Rapid Test Device for the detection of SARS-CoV-2 antigen was compared to RT-qPCR.
RESULTS: 255 nasopharyngeal swabs, including 150 from the emergency department and 105 from primary helthcare centers, were tested. 184 patients were symptomatic (72.1 %). Amongst the 60 positive RT-qPCR samples, 40 were detected by the rapid antigen test, given an overall sensitivity of 73.3 %. All the samples detected positive with the rapid antigen test were also positive with RT-qPCR. The median cycle threshold was 23.28 (IQR 18.5-30.16). Patients with less than seven days onset of symptoms showed a higher viral load, and sensitivity for rapid antigen test (86.5 %), compared to those with more days (sensitivity of 53.8 %)(p < 0.004).
CONCLUSIONS: The rapid antigen test evaluated in this study showed a high sensitivity and specificity in samples obtained during the first week of symptoms and with high viral loads. This assay seems to be an effective strategy for controlling the COVID-19 pandemic for the rapid identification and isolation of SARS-CoV-2 infected patients.