BACKGROUND AND PURPOSE: Concave dose distributions generated by intensity modulated radiotherapy (IMRT) were applied to re-irradiate three patients with pharyngeal cancer.
PATIENTS, MATERIALS AND METHODS: Conventional radiotherapy for oropharyngeal (patients 1 and 3) or nasopharyngeal (patient 2) cancers was followed by relapsing or new tumors in the nasopharynx (patients 1 and 2) and hypopharynx (patient 3). Six non-opposed coplanar intensity modulated beams were generated by combining non-modulated beamparts with intensities (weights) obtained by minimizing a biophysical objective function. Beamparts were delivered by a dynamic MLC (Elekta Oncology Systems, Crawley, UK) forced in step and shoot mode.
RESULTS AND CONCLUSIONS: Median PTV-doses (and ranges) for the three patients were 73 (65-78), 67 (59-72) and 63 (48-68) Gy. Maximum point doses to brain stem and spinal cord were, respectively, 67 Gy (60% of volume below 30 Gy) and 32 Gy (97% below 10 Gy) for patient 1; 60 Gy (69% below 30 Gy) and 34 Gy (92% below 10 Gy) for patient 2 and 21 Gy (96% below 10 Gy) at spinal cord for patient 3. Maximum point doses to the mandible were 69 Gy for patient 1 and 64 Gy for patient 2 with, respectively, 66 and 92% of the volume below 20 Gy. A treatment session, using the dynamic MLC, was finished within a 15-min time slot.
PATIENTS, MATERIALS AND METHODS: Conventional radiotherapy for oropharyngeal (patients 1 and 3) or nasopharyngeal (patient 2) cancers was followed by relapsing or new tumors in the nasopharynx (patients 1 and 2) and hypopharynx (patient 3). Six non-opposed coplanar intensity modulated beams were generated by combining non-modulated beamparts with intensities (weights) obtained by minimizing a biophysical objective function. Beamparts were delivered by a dynamic MLC (Elekta Oncology Systems, Crawley, UK) forced in step and shoot mode.
RESULTS AND CONCLUSIONS: Median PTV-doses (and ranges) for the three patients were 73 (65-78), 67 (59-72) and 63 (48-68) Gy. Maximum point doses to brain stem and spinal cord were, respectively, 67 Gy (60% of volume below 30 Gy) and 32 Gy (97% below 10 Gy) for patient 1; 60 Gy (69% below 30 Gy) and 34 Gy (92% below 10 Gy) for patient 2 and 21 Gy (96% below 10 Gy) at spinal cord for patient 3. Maximum point doses to the mandible were 69 Gy for patient 1 and 64 Gy for patient 2 with, respectively, 66 and 92% of the volume below 20 Gy. A treatment session, using the dynamic MLC, was finished within a 15-min time slot.
Предметные рубрики MeSH
AdultBrain StemCarcinomaCarcinoma, Squamous CellDose Fractionation, RadiationEquipment DesignFollow-Up StudiesHumansHypopharyngeal NeoplasmsMandibleNasopharyngeal NeoplasmsNeoplasm Recurrence, LocalNeoplasm StagingNeoplasms, Second PrimaryOropharyngeal NeoplasmsPharyngeal NeoplasmsRadiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiotherapy, ConformalSpinal CordTime FactorsTreatment Outcome